Use 75 mm Adapter
1
Move the pointer completely to the right. Place the tube adapter in the slot in the center
of the digital reader.
2
Place the spun 75 mm hematocrit tube in tube adapter (clay on the right). Slide the
hematocrit tube completely to the right. Follow steps
in the Read Spun
Hematocrit Tubes inside Rotor to read the tube. Record the result.
Readout Setting
The digital display can be set to read either percent or decimal figures. When the digital
reader is shipped, it is set in the percent mode. The mode can be changed from a percent to
a decimal figure.
1
Remove the base plate on the underside of the digital reader by removing four Phillips
screws. Set the base plate and screws aside.
2
Move the jumper (small black piece) from pins
1 + 2
to pins
2 + 3
.
Figure 2.3
1. Detail A
2. Jumper
3. Percent Mode
4. Pin 3
5. Negative
6. Bulb
7. Positive
8. Fiber Optic Bundle
Operating Instructions
Use 75 mm Adapter
2
55-003823-001CC
2-3