56
CEQ™ 8000 Genetic Analysis System
Table 27: Direct Control Window, Hot Areas
Item
Description
A
Capillary Temperature
-
Used to change the capillary temperature.
B
Gel Capillary Fill
-
Used to fill all capillaries with fresh gel. (Any gel present in the
capillary is forced out to the specified waste position in the buffer plate or wetting tray.)
C
Optical Alignment
- Used to align the lasers with the detection windows of the eight
capillaries. Data can be saved and viewed in the Sequence Analysis module.
D
Manifold Purge
- Used to clear the manifold of air bubbles and/or remaining DNA
fragments before the next separation process.
E
Install/Release Gel Cartridge
- Used to release and/or install a gel cartridge.
F
Denature Samples
- Used to initiate the denaturing of samples.
G
Plate Position
- Used to select the plate or tray position where the capillaries will be
immersed.
H
Wetting Tray
- Used to replace or re-fill the wetting tray.
I
Install/Release Capillary
- Used to install the capillary array. Verify or change the
information requested, and then install the array.
Access Plates
- Displays the
Access Plates
dialog box. This dialog box is used to:
• select the position for capillary immersion in the plate or wetting tray.
• move the plates to the Load position at the front of the instrument to load or unload
plates.
Inject
- Used to initiate the injection of the sample prior to separation.
Separate
- Used to initiate the separation of injected samples.
Summary of Contents for CEQ 8000
Page 42: ...Program Description 28 CEQ 8000 Genetic Analysis System...
Page 98: ...84 CEQ 8000 Genetic Analysis System...
Page 110: ...96 CEQ 8000 Genetic Analysis System...
Page 120: ...106 CEQ 8000 Genetic Analysis System...
Page 128: ...114 CEQ 8000 Genetic Analysis System...
Page 152: ...138 CEQ 8000 Genetic Analysis System Figure 80 Report Format dialog...
Page 154: ...140 CEQ 8000 Genetic Analysis System...
Page 162: ...Run Procedures 148 CEQ 8000 Genetic Analysis System...
Page 220: ...Sequence Analysis Procedures 206 CEQ 8000 Genetic Analysis System...
Page 318: ...Fragment Analysis Procedures 304 CEQ 8000 Genetic Analysis System...
Page 329: ...Exporting Database Items User s Guide 315 Figure 180...
Page 364: ...Direct Control and Replenishment 350 CEQ 8000 Genetic Analysis System...
Page 380: ...Routine Maintenance 366 CEQ 8000 Genetic Analysis System...