Routine Maintenance
User’s Guide
355
Figure 219:Remove Capillary Dialog
15. While holding the Array Fitting tab (of the clean capillary array), align the Array
Fitting with the manifold opening and guide pins. Push the fitting into the manifold
until it is completely seated against the bases of the Guide Pins. See Figure 217.
16. Replace the Manifold Access Cover and tighten the captive screw.
17. Lower the Capillary Temperature Control Cover and secure the two rubber latches.
18. Lower the Capillary Access Cover and Sample Access Cover to their
closed/locked positions.
19. Click
Done
in the Install Capillary Array dialog shown in Figure 220.
Figure 220:Install Capillary Array Dialog