Sequence Analysis Procedures
196
CEQ™ 8000 Genetic Analysis System
Figure 121:Quality Parameters Report
The categories on the Quality Parameters Report are described in Table 70:
Table 70: Quality Parameters Report Categories
Number
the number of the called base
Base
the single letter designation of the called base (A, C, G, T or
other IUB code)
A
base identity score assigned by the base-caller, ranging
from 0 - 255
C
base identity score assigned by the base-caller, ranging
from 0 - 255
G
base identity score assigned by the base-caller, ranging
from 0 - 255
T
base identity score assigned by the base-caller, ranging
from 0 - 255
Call Score
the calculated probability of correctness indicated on a
linear (0.00 - 1.00) scale
Quality Value
the probability of correctness presented on a logarithmic (1
- 100) scale
Migration
the actual migration time (in minutes) in the raw data
corresponding to the called base(s)
Summary of Contents for CEQ 8000
Page 42: ...Program Description 28 CEQ 8000 Genetic Analysis System...
Page 98: ...84 CEQ 8000 Genetic Analysis System...
Page 110: ...96 CEQ 8000 Genetic Analysis System...
Page 120: ...106 CEQ 8000 Genetic Analysis System...
Page 128: ...114 CEQ 8000 Genetic Analysis System...
Page 152: ...138 CEQ 8000 Genetic Analysis System Figure 80 Report Format dialog...
Page 154: ...140 CEQ 8000 Genetic Analysis System...
Page 162: ...Run Procedures 148 CEQ 8000 Genetic Analysis System...
Page 220: ...Sequence Analysis Procedures 206 CEQ 8000 Genetic Analysis System...
Page 318: ...Fragment Analysis Procedures 304 CEQ 8000 Genetic Analysis System...
Page 329: ...Exporting Database Items User s Guide 315 Figure 180...
Page 364: ...Direct Control and Replenishment 350 CEQ 8000 Genetic Analysis System...
Page 380: ...Routine Maintenance 366 CEQ 8000 Genetic Analysis System...