PN B10093AE
A-2
Preinstallation Requirements
Preinstallation Requirements
To ensure optimal safety, the centrifuge should be wired to a remote emergency switch (preferably
outside the room where the centrifuge is housed, or adjacent to the exit from that room). In case of
a malfunction, the centrifuge can be disconnected from the main power source by removing the
Mains (power) plug from the outlet receptacle.
WARNING
To reduce the risk of electrical shock, this equipment uses a three-wire or five-
wire electrical cord and plug to connect the centrifuge to earth-ground. To
preserve this safety feature, make sure that the matching wall outlet receptacle
is properly wired and earth-grounded.
Prior to instrument purchase, the power configuration should be determined and the appropriate
instrument ordered.
Single and Three-Phase Power Connections
shows the power connection for single-phase centrifuges, including earth-ground and
two power leads with 30-ampere circuit breakers.
shows the power connection for three-
phase centrifuges, including earth-ground and three power leads with minimum 16-ampere circuit
breakers and common neutral connection.
contains wiring information.
Table A.1
Nominal Supply Voltage Ratings for the Avanti J-26S XPI
Nominal
Instrument
Voltage Rating
Instrument
Part
Number
Instrument
Part Number
(Elutriation
System)
Nominal Supply
Frequency
Power Cord and Plug Description
Single-phase,
200/208/240 V
B14538
B14544
180–264 V,
50/60 Hz, 30 A
permanently attached three-wire
UL/CSA-approved cord with NEMA
type 6-30P plug
Single-phase,
230 V
B14539
B14545
180–264 V,
50 Hz, 30 A
permanently attached three-wire
CENELEC harmonized cord without
plug
Three-phase,
220/380 V +
Neutral
a
a. Unbalanced three-phase. Split for single-phase operation internally.
B14540
B14546
313–457 V,
plus neutral,
50 Hz, 16 A
permanently attached five-wire
CENELEC harmonized cord without
plug
Summary of Contents for Avanti J-26S XPI
Page 4: ...PN B10093AE iv Revision History ...
Page 12: ...xii Contents Avanti J Series Centrifuge Warranty Related Documents ...
Page 14: ...xiv Illustrations ...
Page 16: ...xvi Tables ...
Page 20: ...PN B10093AE xx Introduction Multi Compliance Label ...
Page 36: ...PN B10093AE 1 16 Description Available Rotors ...
Page 66: ...PN B10093AE 3 10 Troubleshooting JCF Z Rotor Identification ...
Page 82: ...PN B10093AE A 10 Preinstallation Requirements Using J2 Series Rotors in the Avanti J 26S XPI ...
Page 86: ...PN B10093AE Warranty 2 Beckman Coulter Inc Avanti J Series Centrifuge Warranty ...
Page 87: ......