Measurement Accessories
4.4 Measurement Accessories
The photograph below shows a “phantom”. The phantom is
constructed of Perspex material which has known, stable SOS
behavior. It is used to verify that Omnisense is measuring speed
of sound accurately. The procedure is described in Chapter 6,
"System Quality Verification‚" page 6-1.
Figure 4.6
Phantom for system quality verification
A hand rest is provided with Omnisense for use in positioning a
patient’s arm correctly for a SOS measurement of the distal 1/3
radius. The hand rest is illustrated in the photograph below.
Figure 4.7
Limb positioning fixtures
SOS measurements of the distal radius are taken at about the
midpoint of the limb, while measurements of the phalanx and
metatarsal are located along the respective bones. To aid in
finding the correct location, Omnisense is shipped with one or
The phantom has an
LCD temperature gauge
on one side, and a con-
version table on the
other.
The hand rest supports
the lateral edge of the
outstretched palm when
scanning the radius.
Summary of Contents for Sunlight Omnisense 7000S
Page 4: ...iv ...
Page 12: ...xii ...
Page 18: ...xviii ...
Page 22: ...xxii ...
Page 32: ...Omnisense 7000S 8000S 1 10 Conventions Used in this Guide ...
Page 108: ...Omnisense 7000S 8000S 5 14 Changing Users and Shutdown ...
Page 140: ...7 22 User Guide CHAPTER 7 Patient Files ...
Page 162: ...Omnisense 7000S 8000S 8 22 Measuring SOS at the Radius ...
Page 202: ...10 28 User Guide CHAPTER 10 System Administration ...
Page 216: ...Omnisense 7000S 8000S 12 4 Cleaning and Disinfection ...
Page 264: ...Omnisense 7000S 8000S 16 26 System Upgrade ...