Omnisense 7000S/8000S
15-11
Metatarsal V
12.
Scan by moving the probe in a direction perpendicular to the
long mark for a distance of one to two centimeters in the
dorsal direction, and then back again to the starting point.
Repeat this movement continuously until the scan cycle
completes. Do not move the probe quickly.
Figure 15.13 Scanning the metatarsal V
To ensure accurate and reproducible results, note the
following:
•
The probe should be maintained parallel to the metatarsal
bone, never diagonal to it.
•
Never hold the probe at an angle to the surface of the limb.
The probe should be held so that the short dimension of its
surface is tangent to the bone while its long dimension is
parallel to the bone axis.
•
Do not use a circular motion. The probe movement should
follow parallel to the marked line.
•
Probe movement should extend at least 1 centimeter but not
more than 2 centimeters dorsal to the long marked line.
•
Maintain the probe’s surface in contact with the patient’s
skin throughout the scan. No part of the probe surface
should be raised from the skin at any point during the scan.
Hold the probe against
the flat surface of the
bone, and move it in a
range of one to two centi-
meters toward the dorsal
side and back, repeating
until the scan completes.
Summary of Contents for Sunlight Omnisense 7000S
Page 4: ...iv ...
Page 12: ...xii ...
Page 18: ...xviii ...
Page 22: ...xxii ...
Page 32: ...Omnisense 7000S 8000S 1 10 Conventions Used in this Guide ...
Page 108: ...Omnisense 7000S 8000S 5 14 Changing Users and Shutdown ...
Page 140: ...7 22 User Guide CHAPTER 7 Patient Files ...
Page 162: ...Omnisense 7000S 8000S 8 22 Measuring SOS at the Radius ...
Page 202: ...10 28 User Guide CHAPTER 10 System Administration ...
Page 216: ...Omnisense 7000S 8000S 12 4 Cleaning and Disinfection ...
Page 264: ...Omnisense 7000S 8000S 16 26 System Upgrade ...