Connections and Parts 

 

 

 

 

 

Fine Adjust handwheel (clockwise to increase pressure). 

 

Cylinder 

 

Coarse Adjust handwheel (clockwise to increase pressure). 

 

Pressure connector. Default: G 1/8" male with 60° internal cone. 

 

Release valve (clockwise to close/ counterclockwise to open). 

 

 

Summary of Contents for PGL

Page 1: ...INSTRUCTION MANUAL LOW PRESSURE CALIBRATION PUMP MODEL PGL ...

Page 2: ......

Page 3: ...lity of the contents of this manual For more detailed technical data about the Instruction manual for Beamex PGL Low Pressure Calibration Pump please contact the manufacturer Copyright 2013 2017 BEAMEX OY AB Ristisuonraitti 10 FIN 68600 Pietarsaari FINLAND Tel 358 10 5505000 Fax 358 10 5505404 E mail sales beamex com service beamex com Internet http www beamex com 8804200 PGLuEng Version 2 ...

Page 4: ......

Page 5: ...rts 2 Connection Diagram 3 OPERATION 3 Positive Pressure Calibration 3 Negative Pressure Vacuum Calibration 4 Keep in Mind 5 TROUBLESHOOTING MAINTENANCE 6 Seal Replacement 6 Leak Tests 7 Leak Test for Positive Pressure 7 Leak Test for Negative Pressure Vacuum 7 SPECIFICATIONS 8 WARNINGS 9 ...

Page 6: ......

Page 7: ... in PGL o G 6 4 nipple to G 1 8 including an o ring 10 1 mm x 1 6 mm NBR 70 Delivered as a replacement for the pre installed connector A set of seals o rings type 10 1 mm x 1 6 mm NBR 70 Nitrile Butadiene Rubber Shore A70 black Guarantee leaflet This Instruction Manual Optional Accessories The optional accessories are as follows ITEM ORDER CODE Carrying case 8003310 Service Seal kit 8003085 Bag of...

Page 8: ... Adjust handwheel clockwise to increase pressure Cylinder Coarse Adjust handwheel clockwise to increase pressure Pressure connector Default G 1 8 male with 60 internal cone Release valve clockwise to close counterclockwise to open ...

Page 9: ...surement system is not leaking See Leak Tests on page 7 4 Increase the pressure close to the next calibration point by rotating the Coarse Adjust handwheel clockwise 5 Advance to calibration point by rotating the Fine Adjust handwheel 6 Continue to next calibration point by repeating stages 4 and 5 until you reach the calibration point with the highest pressure 7 Decrease the pressure close to the...

Page 10: ...e close to the next calibration point by rotating the Coarse Adjust handwheel counterclockwise 15 Advance to calibration point by rotating the Fine Adjust handwheel 16 Continue to next calibration point by repeating stages 14 and 15 until you reach the calibration point with the deppest vacuum 17 Increase the pressure close to the next calibration point by rotating the Coarse Adjust handwheel cloc...

Page 11: ... variation In that case adjust the pressure back to required value using the Fine Adjust handwheel Do not exceed the max operating pressure of the pump and the hose Also observe the pressure limit set by environmental conditions Ensure that there always is a reliable pressure indicator connected to the pump when the pump is used If there is a strong counterforce while operating the pump but no pre...

Page 12: ...he o ring The thread surface is not smooth Apply Teflon tape The connector types do not match Use proper adapter Hard to Turn the Valves CAUSES SOLUTIONS Release valve closed too tight after the previous calibration Do not close release valve too tightly The new pump does not run smooth The new pump needs time to run in Lack of lubrication in threads Lubricate the thread Seal Replacement Depending...

Page 13: ...00 mbar 6 psi Start your clock After five minutes record the pressure indication p5 After ten minutes record the pressure indication p10 Calculate the leak Lp using this equation 5 10 5 p p p p L If Lp is less than 0 01 1 the system is leak free Leak Test for Negative Pressure Vacuum Decrease pressure using constant change rate down to pump s full scale negative pressure 400 mbar 40 kPa 6 0 psi St...

Page 14: ...o 6 psi Safety Pressure 4 bar 60 psi Pressure media Clean air Temperature 0 50 C Humidity 95 RH Adjusting sensitivity 1 Pa 0 01 mbar 0 004 inH2O 15 10 5 psi Output connectors G 1 8 male with 60 internal cone for Beamex pressure measurement hose pre installed G 6 4 nipple delivered as an optional connector to replace the one meant for Beamex pressure measurement hose ...

Page 15: ...s Use only measuring hoses provided by Beamex or other reliable suppliers Ob serve the effect of the operational conditions to the maximum pressure allowed in the hose The environmental conditions ambient pressure and temperature may restrict the allowable maximum pressure to a lower level than the pump and the hose enable Always depressurize PGL when it is left on its own Avoid possible damage wh...

Page 16: ...ttp www beamex com Beamex Inc 2152 NW Parkway Suite A Marietta GA 30067 U S A Phone 800 888 9892 1 770 951 1927 Fax 1 770 951 1928 E mail beamex inc beamex com Beamex Limited Newtown Grange Farm Business Park Desford Road NEWTOWN UNTHANK Leicestershire LE9 9FL United Kingdom Phone 01455 821 920 Fax 01455 821 923 E mail beamex ltd beamex com Representative ...

Reviews: