62 - Tools
Tool
Description
Available in
Additional
Stability Check
Additional Stability Check is
available for
temperature
measurements. It is an
addition to the default
stability check ensuring that
slow temperature transients
are handled correctly.
For more information see
chapter
• Temperature
Calibrator
• Calibrator
• Documenting
Calibrator*
• Data Logger
*
CMX has fixed default
of 1 min when sending
to calibrator, but if
user changes it in the
calibrator, it will be
received by CMX and
stored with the results.
Additional Info
Including additional
information fields at the
bottom of a subwindow
is always available. The
available fields, however,
depend on the quantity/
settings. Up to four fields
may be added to both
subwindows. The additional
information row settings are
saved for future needs.
• Temperature
Calibrator
• Calibrator
Function Info
Always available. Opens
a pop-up window with
information of current
function (measurement
range, uncertainties etc.).
• Temperature
Calibrator
• Calibrator
• Documenting
Calibrator
• Data Logger
Note:
In
Temperature Calibrator or Calibrator
also the
Additional Information row settings for all functions are saved
for future needs. However Damping, Resolution and Alarms are
active for the current session only.
Additional Stability Check
The
Additional Stability Check
is an enhanced method for determining
when a temperature signal is stable enough for reliable calibration.
The
Additional Stability Check
tool can be activated for the Internal
Reference, RTD and TC sensor measurements. The user can select