background image

 Installation, Operation and Maintenance Instructions

Ohmeda Retro

fi

t Gas Control Panel

Part number 4107 9027 47.00

March 16, 2020

Summary of Contents for 4107 9027 47.00

Page 1: ...Installation Operation and Maintenance Instructions Ohmeda Retrofit Gas Control Panel Part number 4107 9027 47 00 March 16 2020...

Page 2: ......

Page 3: ...l number Option s included Any information service or spare parts requests should include the unit model number and be directed to BeaconMed s 1059 Paragon Way Rock Hill SC 29730 Phone 803 817 5600 Fa...

Page 4: ......

Page 5: ...ts 1 1 4 Environmental Declarations 1 1 5 About Gas Control Panels 1 2 Removal of Old Panel 2 2 1 Remove Ohmeda Front Panel 2 3 Installing New Panel 2 3 1 Installing Retrofit Kit 2 4 Testing 3 4 1 Tes...

Page 6: ......

Page 7: ...ered as well as the environmental perils and dangers to health during the recycling and disposal of the unavoidable rates of non recyclable materials BeaconMed s products for the most part consist of...

Page 8: ...e 1 CAUTION Prior to removal and installation of control panels verify zone valve is shut o and line to control panel is not pressurized Cover Plate Regulator Adjusting Screw and Knob Tubing Connectio...

Page 9: ...Operating Instructions 1 Connect surgical tool supply hose to outlet connection on control panel or remote outlet served by control panel 2 Adjust surgical tool operating pressure turning pressure ad...

Page 10: ...Carefully pull front panel with control assembly out of mounting box Support control assembly in position which allows full view of all connections 3 Adjust control pressure to 160 psig Leak test all...

Page 11: ...5 4107 9027 47 00 Wall Mount Gas Control Panels Notes...

Page 12: ...1059 Paragon Way Rock Hill SC 29730 888 4 MEDGAS 888 463 3427 Fax 803 817 5750 www beaconmedaes com...