9. Pour 2 L of undiluted bleach into an empty cubitainer.
10. Apply a biohazard label to the top of the empty 20-L
cubitainer and label it as WASTE.
11. Load the empty cubitainer into the right side of the
BD FACSFlow supply system unit.
Make sure the biohazard and WASTE labels are facing the
front of the carrier and can be read.
12. Insert the sensor probe into the waste cubitainer, replace the
neck support collar, and tighten the cap assembly.
13. Press Restart.
More information
•
Flushing the BD LSRFortessa system (page 30)
•
Replacing the sheath cubitainer (page 26)
Flushing the BD LSRFortessa system
Introduction
This topic describes the procedure for flushing the BD LSRFortessa
system. Flush the system every two weeks.
About the stainless
steel sheath
containers
The cytometer system is shipped with a stainless steel container for
sheath fluid. This container is replaced by the sheath cubitainer
when the BD FACSFlow supply system is installed. Save the
stainless steel container and tubing and use it when flushing the
system.
Caution!
Replace and tighten all caps when
reassembling the containers and cubitainers. Loose
caps can result in fluid leakage.