66
BD FACSCanto II Flow Cytometer Reference Manual
Unscheduled Maintenance
Perform these maintenance procedures as indicated
in the table.
Procedure
Description and Location
When to Perform
Replace cubitainers
See Changing a Cubitainer on
page 68.
As needed
Prime fluid lines
Removes air from fluid lines
See Priming Fluidics Lines on
page 71.
Whenever a fluidics line
is disconnected to change
a cubitainer or perform
other maintenance
Remove an air lock
Removes air from filter and
restores flow of fluid
See Removing an Air Lock on
page 72.
When fluidics are not
functioning properly
(flow cell will not fill, or
there is backflush into
sample tube)
Clean external surfaces
Keeps surfaces free from salt
buildup after exposure to sheath
fluid
See Cleaning External Surfaces
on page 74.
As needed
De-gas Flow Cell
Removes bubbles from the flow
cell
See See Removing Bubbles from
the Flow Cell on page 76.
As needed during daily
startup
Clean Flow Cell
Runs BD FACSClean solution
through the SIT and flow cell
See Cleaning the Flow Cell on
page 76.
When poor optical
performance indicates
additional cleaning is
needed
Bubble Filter Purge
Removes air from the bubble
filter
See Purging the Bubble Filter on
page 78.
If fluidics run dry, or
when CVs are poor
Summary of Contents for FACSCanto II
Page 4: ......
Page 12: ...xii BD FACSCanto II Flow Cytometer Reference Manual...
Page 44: ...44 BD FACSCanto II Flow Cytometer Reference Manual...
Page 104: ...104 BD FACSCanto II Flow Cytometer Reference Manual...
Page 124: ...124 BD FACSCanto II Flow Cytometer Reference Manual...
Page 138: ...138 BD FACSCanto II Flow Cytometer Reference Manual...
Page 144: ...144 BD FACSCanto II Flow Cytometer Reference Manual...
Page 152: ...152 BD FACSCanto II Flow Cytometer Reference Manual...