     •  Time that past history was   

       LAST CLEARED

4.  Press the 

ZOOM

 key 

(time interval)

 as appropriate and  

review drug totals.

5.  To clear the patient history, press 

CLEAR HISTORY

 and  

then press 

YES

 or 

NO.

6.  To view 24-hour totals: Press 

24 H Totals.

 

Note:

 This is a moving window of time, meaning that  

after 24 hours the patient history rewrites.

7. Press 

EXIT

 and then press 

START.

PCA/Monitoring trend data:
Note:

 This function requires use of the Alaris PCA  

monitoring module(s):

1.   Press 

CHANNEL SELECT

 on 

monitoring

 module.

2.   Press 

OPTIONS

.

3.   Press 

PCA/Monitoring trend data

.

4.   To exit: Press 

(EtCO

2

 or SpO

2

)

 

MAIN

.

5.   Press 

MAIN SCREEN

.

Change PCA Pause alarm limits:

1.   Press the 

CHANNEL SELECT

 key on the PCA module.

2.   Press 

OPTIONS

.

3.   Press 

PCA PAUSE LIMITS

.

4.

   Select desired parameter and enter value. 

 

Note:

 If acceptable range value is 

not

 within the  

 hospital defined range, a prompt is provided.

5.   Follow on-screen prompts.
6.   Press 

CONFIRM

 and press 

START

.

Responding to a PCA Pause alarm:

1.   Always follow hospital protocol.

2.  

Press 

CONFIRM.

  

Note:

  Main screen shows PAUSED. The Alaris PCA module 

cannot be restarted until the patient’s monitoring values have  
been reestablished and are within hospital established limits.

3. 

Press 

RESTART.

 

Note:

 To view time and patient PCA pause value  

that caused the PCA module to pause, access drug  
event history.

Disabling and enabling the PCA Pause alarm:

1.  Press the 

CHANNEL SELECT

 key on the PCA module.

2.  Press 

OPTIONS

.

3.  Press 

PCA PAUSE LIMITS

4.  Select 

Disable SpO

2

 or 

Disable EtCO

2

.

5.  Press 

CONFIRM

.

6.  Press 

START

.

Note:

 To enable the PCA Pause function, follow steps 1–3,  

then press 

Enable SpO

2

 or 

Enable EtCO

2

.

Change dose request cord audio and light setting:

1.  Press the 

CHANNEL SELECT

 key.

2.  Press 

OPTIONS

.

3.  Press 

Dose Request Setup

.

4  

Select desired dose cord profile. 

(1=light flashes, 2=light on, 3=light off)

5.  Press 

CONFIRM.

 Then press 

START

.

To attach the dose request cord:

Align the red markings on both the Alaris PCA module and  
Dose Request Cord.

To detach the dose request cord:

Hold the body of the latching connector on the Dose Request  
Cord and pull straight away from the Alaris PCA module,  
without twisting or turning. 

Detaching the Alaris PCA module:

Use the PCA key to unlock the door. Locate the black lever inside  
at the bottom left and depress. At the same time: Hold the Alaris  
PCA module and move the bottom of the Alaris PCA module  
away from the Alaris PC Unit.

Troubleshooting:

 

Alarms, errors, messages

Note:

 During an alarm state, the Alaris PCA module is 

not

 infusing.

Incorrect concentration or dosing:

An incorrect concentration or dosing parameter may have  
been programmed, which could result in an excessive volume  
or dose being delivered.  

1.  Remove the syringe.

2. 

Verify the concentration listed on the syringe matches the 

concentration 

(DRUG AMOUNT and DILUENT VOLUME)

 

programmed into the PCA module. 

3.  Reprogram.

Note:

 This message can be the result of an incorrect DRUG 

AMOUNT and/or DILUENT VOLUME entry, or it can occur if  
hospital-established Guardrails

TM

 software limits are very wide.   

Be sure to enter either a drug amount per 1 mL or total drug 
amount per total volume. For example, a 30 mL syringe with 
concentration of 1 mg/1 mL can be entered in one of two ways:

•  DRUG AMOUNT 1 mg
•  DILUENT VOLUME 1 mL  
    or
•  DRUG AMOUNT 30 mg
•  DILUENT VOLUME 30 mL

Maximum limit reached:

The programmed maximum limit has been reached over the  
time period specified. Infusion paused until time limit has expired. 
Alarm message 

Max Limit Reached

 will scroll in channel message 

display on the Alaris PCA module. To silence safety alarm tone, 
press the 

SILENCE

 key. The Alaris PCA module will remain silent.

Near End of infusion alert (NEOI):

The alert message, 

Near End

 alternates with remaining 

VTBI

 on 

the screen until the syringe is empty. The alert message will scroll 
in the channel message display on the Alaris PCA module. The 
Alaris PCA module remains functional and will continue infusing. 
To silence the safety alert tone, press the 

SILENCE

 key. The Alaris 

PCA module will remain silent until the 

Syringe Empty

 alarm 

sounds. When programmed in PCA dose-only mode, the green 
indicator light illuminates only when a PCA dose is being delivered. 
When programmed in PCA dose and continuous mode, the green 
indicator light remains illuminated. The yellow indicator light will 
flash during an NEOI alert.

Syringe Empty:

The alarm message, 

Syringe Empty

 will scroll in the channel 

message display on the Alaris PCA module. To silence the safety 
alarm tone, press the 

SILENCE

 key. The Alaris PCA module will 

remain silent for approximately two minutes and will sound again.

Patient history/24-hour history:

1.  Press the 

CHANNEL SELECT

 key.

2. Press 

OPTIONS.

3. Press 

PATIENT HISTORY.

 

Note:

 The top of the 

PATIENT  

HISTORY

 screen displays  

   •  The exact time frame for  
       patient history. History from  
       ##:## - ##:##

Reviews: