An alarm indicates that the infusion has stopped. 

Check the display for an alarm symbol. 

Press 

c

 to silence the sound. (Exceptions are 

M

 and 

K

 alarms, which cannot be 

silenced)

When the cause has been rectified, press 

d

 key to resume infusion.

All alarms will stop the infusion except for 

F

 and 

H

where the infusion status will remain the same state as prior to  

the alarm.

Alarm / Condition

Action

J

Air-in-line

Upstream occlusion

• Disconnect the infusion set from the 

IV access device and use the prime 

function.

• Remove occlusion/air and restart the 

infusion.

K

Internal battery depleted.

• Connect pump to AC power.
• Ensure pump is switched on to charge 

internal battery.

L

Door open

• Close door and restart infusion.

M

System fault

• Switch pump off.
• Remove pump from service and have 

the pump inspected by a qualified 

service engineer.

N

Flow error 

Flow sensor attached to secondary 

infusion set.

• Close in-line clamp.
• Ensure tubing is loaded correctly. 
• Ensure ample fluid is in container. 
• Clear blockage. Close pump door. 
• Restart infusion.
• Ensure flow sensor is attached to 

primary infusion set.

O

Flow sensor connection error.

• Connect flow sensor to pump and 

restart infusion. Ensure fluid in drip 

chamber is not above fill line.

P

Downstream occlusion

• Remove pressure in the infusion set to 

prevent a post occlusion bolus to the 

patient.

• Remove the cause of the blockage. 
• Restart the infusion.

Q

Incorrect infusion set, infusion set 

incorrectly loaded or infusion set worn.

• Use infusion set as specified in DFU. 

Ensure infusion set is loaded correctly 

into pump.

Alarms

Summary of Contents for Alaris 800TIG2GBD1

Page 1: ...Alaris GW 800 Volumetric Pump Models 800TIG2GBD1 800TIG2RWN1 Quick Reference Guide en ...

Page 2: ...plete pump information refer to the Alaris GW 800 Volumetric Pump Directions For Use All persons operating the pump should consult the Directions For Use PRESSURE SENSOR TUBING guide DOOR latch AIR SENSOR FLOW STOP DISPLAY INDICATORS MAIN DISPLAY ENTER CHEVRONS RUN HOLD ON OFF PRIMARY SECONDARY PRIME BOLUS CLEAR MUTE MUTE INDICATOR ALARM INDICATOR Controls ...

Page 3: ...te To re enable the alarm audio press any other button or key to unmute PRIME BOLUS button Primes the Infusion set Administers bolus during the infusion PRIMARY SECONDARY button Switches the Pump between Primary and Secondary infusion modes if enabled ENTER button Scrolls between rate time VTBI and total volume infused VI Enters values for selected infusion configuration parameters Confirms the ra...

Page 4: ...the Volume To Be Infused VTBI in millilitres ml VI When illuminated the Pump is displaying the Volume Infused VI in millilitres ml TIME When illuminated the Pump is displaying the infusion time in hours mins MICRO When illuminated the Pump is operating in the Micro Mode When not illuminated the Pump is in the Standard Mode SEC When illuminated the Pump is operating in the Secondary Mode When not i...

Page 5: ...oor 3 Release flow stop mechanism by pushing the lever arm up and to the right 4 Insert the infusion set from left to right into the pumping channel Avoid any slack and avoid stretching or pulling the tubing 5 Engage the flow stop mechanism by pushing the lever left and down 6 Close door Press recesses firmly as shown to ensure latch clicks into place Loading the Infusion Set ...

Page 6: ... visible air is cleared from the line 3 Connect the infusion set to the patient access device Enter rate using h f Press e once to confirm rate and scroll to VTBI 4 Enter VTBI using h f or if using a flow sensor VTBI can be switched off by scrolling until OFF is displayed Press e once to confirm VTBI and scroll to VI 5 If necessary press c to clear VI 6 Press d to start the infusion Starting the P...

Page 7: ... to high HI normal nor or low lo using h f Press e to confirm and go to next option Selecting the Alarm Volume if Enabled tonE option is displayed t Change alarm volume between 1 and 7 using h f Press e to confirm and go to next option 1 Low 7 High Selecting VTBI Time Infusion if Enabled CLoc option is displayed q Select On or OFF using h f keys to turn the VTBI time infusion setting on or off Pre...

Page 8: ...ration To change the rate while infusing if titration enabled 1 Press h f to adjust rate 2 Press e to confirm change of rate Administer BOLUS Press b To administer a bolus while infusing Press b While boL is displayed press b again and hold to administer the desired bolus volume Features ...

Page 9: ... L Door open Close door and restart infusion M System fault Switch pump off Remove pump from service and have the pump inspected by a qualified service engineer N Flow error Flow sensor attached to secondary infusion set Close in line clamp Ensure tubing is loaded correctly Ensure ample fluid is in container Clear blockage Close pump door Restart infusion Ensure flow sensor is attached to primary ...

Page 10: ...ed pre set volume to be infused with KVO switched off Set new VTBI F Low battery At least 30 mins before K alarm Connect pump to AC power source H The pump has been left unattended for two minutes and infusion has not started Attend to pump Alarms continued ...

Page 11: ...key to continue the infusion once prescribed bolus has been administered D VTBI limit reached Pump will infuse at the KVO rate until d is pressed E The pump is priming the infusion set Ensure all air has been primed out of the infusion set before starting the infusion G The pump is on hold Press d to restart continue infusion or press e to return to setup I Automatic infusion set check Allow test ...

Page 12: ... BD Logo and Alaris are trademarks of Becton Dickinson and Company or its affiliates 2019 BD All rights reserved BD Switzerland Sàrl Route de Crassier 17 Business Park Terre Bonne Batiment A4 1262 Eysins Switzerland ...

Reviews: