Change of the display at different range of pressure sensor 
(regardless of the measuring range of the pressure sensor, the pressure reference is necessary) 

During the lifetime of the pressure sensor, the value of the range being nominally at 20,000 mA e.g., might be 
shifted. This phenomenon could cause that the display shows a signal value which differs from the adjusted 
boundary of the measuring range. In the control software of the display, this phenomenon was considered and 
a function eliminating this phenomenon was implemented. 

 

The PAof menu should be selected. 

 

Both the buttons should simultaneously be pressed. 

 

The number 0238 should be entered to select the special function. 

 

Both the buttons should be pressed simultaneously again. The following sample value appears on the 
display. 

 

 

Now, the pressure sensor has to be pressurized via the pressure reference. The pressure has to 
correspond to the ending value of the measuring range.  

 

When pressing both the buttons simultaneously afterwards, the current output signal of the sensor is 
saved as a voltage signal.  From this moment, the adjusted ending of the measuring range (end-point) 
appears on the display, although the sensor signal is shifted in the voltage signal. 

 ATTENTION: Please note that the output signal remains unaffected by this change. 

Entering the data of the basic adjustment of the mechanism ("Load Defaults") 

The control system of the display is equipped with the possibility to re-create the adjustment of the mechanism. 
That way pre-made changes related to the compensation of the offset or the voltage can be performed 
retroactively. 

 

To load the basic adjustment, the PAof menu should be selected. 

 

Then, both the buttons should simultaneously be pressed. 

 

The number 0729 should be entered to select the special function. 

 

Both the buttons should be pressed simultaneously again. The following sample value appears on the 
display, 

 

 

If both the buttons are simultaneously pressed again, the basic adjustment of the mechanism is 
activated again. 

Restrictions on entering a password 

Dealing with special functions for the compensation of the offset, the range, recording the basic adjustment and 
password change, various codes have been entered to call the function.  

 ATTENTION: Please note that these codes cannot be used as passwords. 

 
Structure of the menu system 

 

Summary of Contents for PCB+

Page 1: ...0 C 0 C to 60 C Compensated range of temperatures 0 C to 60 C 10 C to 40 C Influence of temperature in the compensated range of temperatures a zero 0 1 1 8 of range 10 C depends on the pressure range and the sensor type b range 0 1 1 8 of range 10 C depends on the pressure range and the sensor type Material of the box ABS Fig 1 Mechanical dimensions Fig 2 Location of terminal blocks and adjusting ...

Page 2: ...adjusted settings Control elements for setting the functions For the control two buttons located under a foil can be used The user moves forward in the menu system or increases the values on the display by the button The button is used to decrease the values and to move backward in the menu system The menu system is closed By scrolling the particular menu settings forward and backward the desired ...

Page 3: ... signal From this moment the adjusted ending of the measuring range end point appears on the display although the sensor signal is shifted in the voltage signal ATTENTION Please note that the output signal remains unaffected by this change Entering the data of the basic adjustment of the mechanism Load Defaults The control system of the display is equipped with the possibility to re create the adj...

Page 4: ...about exceeding the range After pressing both the buttons simultaneously the report about exceeding or failing to meet the display range can be activated Only the ON status or the OFF status can be selected Pressing both the buttons simultaneously the settings are closed High Pressure After pressing both the buttons simultaneously the maximum pressure saved during the measurement is displayed If b...

Reviews: