BD Sensors DM01 Operating Manual Download Page 2

7.3 Menu description 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

8. Maintenance 

 

DANGER 

Danger of death from airborne parts, 
leaking fluids, electro shock 

Always service the device in a 
depressurized and de-energized 
condition! 

 

WARNING 

Danger of injury from aggressive 
media or pollutants 

-  Depending on the measured medium, 

this may constitute a danger to the 
operator.  

-  Wear suitable protective clothing 

e.g. gloves, goggles. 

If necessary, clean the housing of the device using a moist cloth 
and a non-aggressive cleaning solution. The cleaning medium for 
the media wetted parts (pressure port / diaphragm / seal) may be 
gases or liquids which are compatible with the selected materials. 
Also observe the permissible temperature range according to the 
data  sheet.  Deposits  or  contamination  may  occur  on  the 
diaphragm / pressure port in case of certain media. Depending on 
the quality of the process, suitable maintenance intervals must be 
specified by the operator. As part of this, regular checks must be 
carried  out  regarding  corrosion,  damage  to  the  diaphragm  and 
signal  shift.  If  the  diaphragm  is  calcified,  it  is  recommended  to 
send the device to BD SENSORS for decalcification.  

NOTE  -  Wrong  cleaning  or  improper  touch  may  cause  an 
irreparable  damage  on  the  diaphragm.  Therefore,  never  use 
pointed objects or pressured air for cleaning the diaphragm. 

9. Placing out of service 

 

DANGER 

Danger of death from airborne parts, 
leaking fluids 

Disassemble the device in a  
depressurized and switched-off 
condition! 

 

WARNING 

Danger of injury from aggressive 
media or pollutants 

-  Depending on the measured medium, 

this may constitute a danger to the 
operator.  

-  Wear suitable protective clothing 

e.g. gloves, goggles. 

10. Service / repair 

Information on service / repair: 

www.bdsensors.cz  

[email protected] 

Service phone:  +420 572 411 011

 

10.1 Recalibration 

During the life-time of a transmitter, the value of offset and span 
may shift. As a consequence, a deviating signal value in 
reference to the nominal pressure range starting point or end 
point may be transmitted. If one of these two phenomena occurs 
after prolonged use, a recalibration is recommended to ensure 
furthermore high accuracy. 

10.2 Return 

 

WARNING 

Danger of injury from aggressive 
media or pollutants 

-  Depending on the measured medium, 

this may constitute a danger to the 
operator.  

-  Wear suitable protective clothing 

e.g. gloves, goggles. 

Before every return of your device, whether for recalibration, 
decalcification, modifications or repair, it has to be cleaned 
carefully and packed shatter-proofed. You have to enclose a 
notice of return with detailed defect description when sending 
the device. If your device came in contact with harmful 
substances, a declaration of decontamination is additionally 
required.  
Appropriate forms can be downloaded from our homepage. 
Download these by accessing www.bdsensors.com or request 
them: 
[email protected] | phone: +420 572 411 011 

In case of doubt regarding the fluid used, devices without a 
declaration of decontamination will only be examined after 
receipt of an appropriate declaration! 

11. Disposal 

 

WARNING 

Danger of injury from aggressive 
media or pollutants 

-  Depending on the measured medium, 

this may constitute a danger to the 
operator.  

-  Wear suitable protective clothing 

e.g. gloves, goggles. 

The device must be disposed of according to the 
European Directive 2012/19/EU (waste electrical 
and electronic equipment). Waste equipment must 
not be disposed of in household waste! 

NOTE 

Dispose of the device properly! 

12. Warranty terms 

The warranty terms are subject to the legal warranty period of 24 
months, valid from the date of delivery. If the device is used 
improperly, modified or damaged, we will rule out any warranty 
claim. A damaged diaphragm will not be accepted as a warranty 
case. Likewise, there shall be no entitlement to services or parts 
provided under warranty if the defects have arisen due to normal 
wear and tear. 

13. EU declaration of conformity / CE 

The  delivered  device  fulfils  all  legal  requirements.  The  applied 
directives, harmonised standards and documents are listed in the 
EC  declaration  of  conformity,  which  is  available  online  at: 
http://www.bdsensors.com. 
Additionally, the operational safety is confirmed by the CE sign on 
the manufacturing label. 

 

Switching on 

Switching on without status message

, with "Left button" and "Right button" key possible. 

Switching on with status message

, only possible with button "Menu key" (middle button). 

Status message (appears in the display for approx. 2 seconds): 
-Memory usage: in percent 
-TAG: Measuring point designation in text form 
-Battery: Status of the battery charge 
-Firmware: installed version 

Menu 1/11 
Access 

Password

: **** (a four-digit, freely combinable statement consisting of numbers, letters and special characters) 

-Protection [Off]: unrestricted operation 
-Protection [On]: operation only possible after password input 
(Select menu item "Password" with "Edit" → Press "<<" or ">>" → Set value → continue with "Next". Set password and remember! 
→ Press "Next" to "Protection" sub-item → Press "<<" or ">>" → Activate protection [On] or deactivate protection [Select] → confirm with "Next" and continue to menu bar.)        

 

NOTE 

no connection to the evaluation software BD│DAQ, if password is active! If you have forgotten your password, contact the manufacturer! 

Menu 2/11 
Min/Max Values 

Display of min / max values 

P

min

 - Minimum pressure display:  The minimum pressure applied during measuring is shown in the display. 

P

max

 - Maximum pressure display: The maximum pressure applied during measuring is shown in the display. 

T

min

 - Minimum temperature display: The minimum temperature during measuring is shown in the display. 

T

max

 - Maximum temperature display: The maximum pressure applied during measuring is shown in the display.  

Possible options: reset value [Reset ?, Sure?]  
(Resetting of a value: select the menu point with "Edit" → button ">>" operate. There appears the question "Reset?" → once more operate the button ">>". It seems "Sure?" 
additional confirmation whether the value should be put back → repeated confirming with the button ">>" takes over topically adjoining pressure as a minimum value.) 

Menu 3/11 
Data Logger 

 

Data Logger configuration 

the following settings are possible: linearly [Linear] (value admission to the counter level 600798 is reached), cyclically ( [Loop] (after the value is reached in 600798, the 
data logger automatically  begins the values once more to grasp and, besides, headlines the old values) or [Off] (in the display appears "

D

",  if the data logger is activated 

and goes out if the data logger is off).  
  
Intervals to the memory of the measuring values (pressure / temperature):  
Interval: second [1-99 sec.]; minute [1-99 min]; hour [1-99 h]; or day [1-99 days], the time of day is to be set additionally;  
Milliseconds [20 msec.], only possible if the sampling rate is set to 50 / sec. in menu 4/11 (measuring performance). 
 
Time of day: Measured value recording: at what time the value should be recorded (only effective for the interval setting "day"). 
TAG: Measuring point inscription, factory set BD│Sensors. The setting can be changed by the user. 

NOTE 

- While the data logger is active, the display and pressure sensor module may not be disconnected!

 

Menu 4/11 
Measuring Perf. 

Sample rate: Possible settings [1 / sec.], [2 / sec.] or [50 / sec.] only if the interval is set to [20 msec.] in menu 3/11 (Data Logger). 
Damping: Damping can be set in one-second increments between [1 sec.] and [10 sec.], or disabled by selecting [Off]. 

Menu 5/11 
Quick access 

Button configuration: Left button / Right button 
Left / Right button: configuration of functions: [Min], [Max], [Light], [Zero], [Reset], [Single], [Off] 
 

 

Description of the functions: 

- [Min] / [Max]   minimum / maximum pressure value is shown in the display 
- [Light] 

 The backlight will turn on only when the illumination time in the 8/11 menu is set to 1-10 s.

 

- [Zero] 

 the zero point is set automatically, the display shows "

Z

- [Reset] 

 the set zero point is reset, goes out  

- [Single]             the measured values are recorded individually after pressing the button 
- [Off]                  switches off the display (standby), provided the data logger is deactivated. 

Menu 6/11 
Units 
 

Adjustment of pressure unit 

adjustable units: [bar], [PSI], [mbar], [mH2O], [inHg], [cmHg], [mmHg], [hPa], [kPa], [MPa], [kg/cm2], [ inH2O], [mmH2O] or [User] (the user-defined unit [User] can only be 
programmed using the software BD | DAQ), all pressure-related parameters are converted  

Setting the decimal places 

settable decimal places: standard [Std], one decimal place [+1] or two decimal places [+2] 

Setting the temperature unit 

adjustable units: degrees Celsius [°C], degrees Fahrenheit [°F] or Kelvin [K] set (factory setting [°C]) 

Menu 7/11 
Settings 1 

Setting the date, time and language  

Adjustable options: The date in the format [T.M.JJJJ], the time in the format [hh: mm] and the language [German] or [English]. 

Menu 8/11 
Settings 2 

 

Setting the switch-off time, the lighting and the brightness 

Off time: Setting the automatic switch-off in minutes. The automatic shut-off can be configured in increments of [1 min], [2 min], [3 min], [4 min] or [5 min] (the timer is 
activated 30 sec. before switching it off) or disabled by the [Off] option. After deactivation, the precision digital pressure gauge is in continuous operation. 
Illumination: the illumination duration can be set in one-second increments between [1 s] and [10 s] and in ten-second increments between [20 s] and [120 s], or disabled by 
selecting [Off] and enabling [On]. Note: For continuous lighting [On] increased consumption of the battery charge. 
Brightness: The brightness can be adjusted in 10% increments between [0%] and [100%]. 

Menu 9/11 
Sensor Data

 

Overview of sensor data (pressure sensor module) 

[SN:]  

Serial number (ten-digit number)                                            

[Lower:]         Start of measuring range (value and unit) 
[Upper:] 

Measuring range end (value and unit) 

 

 

[Date:] 

Date of manufacture (dd.mm.yyyy) 

The values are set by the factory and cannot be changed. Automatic detection after connecting the sensor to the display 

Menu 10/11 
Device Info 

Overview of device information (display) 

[SN:] Serial number (eight-digit number) 
[Cap:] Data logger capacity (occupied range 0-600798 / maximum acceptance 600798) 
[Firmware:] The installed firmware version is displayed. 
[Production:] Date of Manufacture (TT.MM.JJJJ) 
Note: The values are set by the factory and cannot be changed. The recorded value in the data logger can be reset. 
 
(Reset counter reading: menu point [Cap:] with "Edit" select →  button "<<" or ">>" press. There appears the question "Reset?"  → once more operate the button "<<" or 
">>". It seems "Sure?" additional confirmation whether the value should be reset → repeated confirming with the button "<<" or ">>" reset the grasped measuring values. 
Display announcement "Counter: 0/600798") 

Menu 11/11 
Service 

Setting the service options 

 

Device restart: [No] or [Yes] Switching off and switching on the device is carried out automatically. Required before firmware upgrade. 
Presets: Reset [No] or [Yes] to factory defaults 

Error 

Display "No sensor": Display and pressure sensor modules are disconnected.  
Indication "Inappropriate sensor": Sensor is not suitable for the sampling rate 50 / s and the interval of 20 ms. 

Firmware update 

Download current firmware (https://www.bdsensors.cz), switch on the device, connect the display to the computer (see 5.2), start the firmware update tool. Device (display)  
restart (automatic detection), select update file with Select File, press Start Update button and execute update. Important: the update may not be interrupted! 

Left button

:     is a function button and can be configured in menu 5. Off, Min, Max, Light, Zero, Reset or Single function can be assigned to the button. The configured function is 

active in display mode. Hold the button for about 2 seconds to activate the preset function. In operating mode, move backwards in the menu system "<<" or reduce 
the setting value. 

 

Right button:

   is a function key and can be configured in menu 5. Off, Min, Max, Light, Zero, Reset or Single functions can be assigned to the key. Hold the button for about 2 seconds

 

                          to activate the preset function. In operating mode, move forward in the menu system ">>" or increase the setting.

  

 

Menu-button:

   pressing this "Menu" button will enter the operating mode; It also serves to select the individual menu items "Edit" or to confirm the set values "Next".

 

                          When pressing the button for approx. 4 seconds, the operating mode is exited.

  

 

                                                                                                                                                                        
To configure the individual menu items, the desired menu item must be set with the help of the left key "<<" or the right key ">>". Then confirm this with the menu button "Edit". Menu item 
is highlighted and configuration can begin. 
To save a set value the menu key "Next" must be pressed. To exit the menu, press the menu button for approx. 4 seconds. The operating mode is also left automatically after approx. 1 
min. 
 

 

Changes are only effective after pressing the menu button "Next" and after leaving the menu item. When leaving the entire menu system, the set parameters are checked 
again in relation to each other and in relation to the characteristics of the device. When configuring the unit, the measuring range is converted into the new unit only after leaving 
the menu system. Depending on the pressure range, not all units may be used.

 

 
 
 

Notes 

 

_____________________________________________________________________________________________
_____________________________________________________________________________________________
_____________________________________________________________________________________________
_____________________________________________________________________________________________
_____________________________________________________________________________________________
_____________________________________________________________________________________________
_____________________________________________________________________________________________
_____________________________________________________________________________________________
_____________________________________________________________________________________________
_____________________________________________________________________________________________
_____________________________________________________________________________________________
_____________________________________________________________________________________________
_____________________________________________________________________________________________
_____________________________________________________________________________________________
_____________________________________________________________________________________________
_____________________________________________________________________________________________
_____________________________________________________________________________________________
_____________________________________________________________________________________________
_____________________________________________________________________________________________
_____________________________________________________________________________________________
_____________________________________________________________________________________________
_____________________________________________________________________________________________
_____________________________________________________________________________________________
_____________________________________________________________________________________________

Reviews: