background image

TM

TM

4

5

Once the desired option is visible, pressing the Power Button will select that option.
Note: The scanner will store the last 300 frames (10-12.5 seconds depending on the selected 

scan depth) in its internal memory. To store on the scanner, the scanner must be Frozen. During 

save, particularly cine save, the app will pause while frames are downloading and saving the data 

while scanning. It is recommended that you save images and videos directly from the app where 

possible. 

2 LED sequence

Freeze the image using 

“Power button”

 

Press

 “Down”

 will bring up “

Save Image

”.

Freeze the image using 

“Power button”

 

Press “

Up

” will bring up “

Save Cine

”.

1.2 Saving images and videos using the scanner buttons

Power 

LED

Wi-Fi

LED

3.2 Normal start up process

1. Press “

Power

button

” to switch 

on the scanner. 

Open app on the 

smart device. The 

app will request 

to connect to the 

scanner. Click 

OK.

3 App requests 

to connect to 

the internet. 

Click OK.

4. Select your 

local Wi-Fi 

network to 

connect to the 

internet. 

5. App retrieves 

the license 

from BCF’s 

license server 

and the app 

will request to 

connect back to 

the scanner. 

2. Select 

EASISCAN 

from the Wi-Fi 

page and 

return to the 

app.
The app will 

get the serial 

number from 

the scanner.

6. Select your 

EASISCAN from 

the Wi-Fi page 

and return to 

the app.

3 Activation

3.1 First time use

Download the BCF Go Scan app           from Google Play or the App store.
You are likely to be prompted for a scanner software update on first time use. Please see 

Scanner Update section.  
To activate the scanner, ensure that the scanner is charged and that you have a local Wi-Fi 

internet connection. This is only necessary the first time you connect the scanner.

After a few seconds, the ultrasound image will appear. Touch anywhere on the screen to bring up 

the live scanning menu.

Note - if the Wi-Fi LED is flashing on the scanner it is not connected to any smart device.

1. Switch on the 

scanner and wait 

until the Wi-Fi 

LED illuminates. 

Launch BCF 

Go Scan App. 

App requests 

to connect to 

scanner.

2. Select your 

EASISCAN 

from the Wi-Fi 

page. Return to 

the App.

3.It can take up 

to 10 seconds 

to connect and 

start receiving 

ultrasound. You 

are now ready 

to scan. 

Power LED

Blue 

flashing

Scanner turning on

Blue

Scanner on and battery 

between 66–100% charge

Yellow

Scanner on and battery 

between 33–66% charge

Yellow 

flashing

Scanner on and battery 

between 0–33% charge

Wi-Fi LED

Blue 

flashing

Scanner ready for smart 

device to connect, nothing 

connected

Blue

Smart device connected

Both LEDs

Flashing 

yellow

Software updating

Both LEDs

Flashing 

yellow

Software updating

3.3 Scanner Updates

 

The app automatically checks the BCF licence 

server for scanner software updates when 

connected to Wi-Fi.
After connecting to an EASISCAN, you may 

be prompted to update the scanner software. 

The scanner will only update if the battery is 

above 50% charge.
Press OK or Cancel. Pressing OK will cause 

the scanner LEDs to flash yellow - the 

scanner will take roughly two minutes to 

update. Press cancel if you wish to update 

later. 
Once the update is finished the scanner will 

switch off. The scanner software version 

number can be checked in the User Settings 

section.

Summary of Contents for Easi-Scan:Go

Page 1: ...TM Your animal imaging partner Instruction manual...

Page 2: ...erence and 2 This device must accept any interference including interference that may cause undesired operation of the device Cet quipement est conforme et se soustrait de Licence au Cahier des Charge...

Page 3: ...Update section To activate the scanner ensure that the scanner is charged and that you have a local Wi Fi internet connection This is only necessary the first time you connect the scanner After a few...

Page 4: ...eeze the image Change the channel in Wi Fi settings No indicator clear channel on the 5GHz band This is recommended state White indicator clear channel but is on 2 4GHz band it is more prone to interf...

Page 5: ...width of the skull at the widest point 14mm 101mm 49 126 days 19mm 66mm 48 121 days Gestational measurements available Trunk diameter width of the ribcage at the widest part 14mm 123mm 53 133 days Ey...

Page 6: ...ea will return to live scanning 4 3 Settings Freedraw Touch and drag to draw on screen can be used to make simple annotations or to highlight areas of interest Any measurements placed on screen remain...

Page 7: ...and unfreeze Texts from Visits ToDo App Entering and exiting standby mode Files can be uploaded to BCF s license server www licencing bcftechnology com Automatically if launching app when connected t...

Page 8: ...e unit is NOT corrosion proof and thus it should not be left wet Take care to detach the external battery and ensure interconnect areas are dry also The battery charger and charging carry case are not...

Page 9: ...AD 12VDC 2A Fused Only to be used when charging from a vehicle supply Only replace fuse with 2A medium acting fuse rated to 125V 10kA breaking capacity F2A M125V DC power Adapter XP Power Model VEP24U...

Page 10: ...fficial authorised service center or the distributor in your country In the UK BCF Technology Ltd Imaging House Phoenix Crescent Strathclyde Business Park Bellshill ML4 3NJ Scotland UK T 44 0 1506 460...

Page 11: ...TM 20 BCF Technology Ltd V26 May 2018 Easi Scan Go...

Reviews: