ULT-2000 Series Quick Start Guide (SonoSite Ultrasound)
–
10/2019
If using a soak tube* (pictured, left):
-
Insert the tip of the ULT-PC-30 Submersible Conductivity Probe to the bottom of the tube.
-
Fill the tube to an appropriate level with cleaning solution, saline, or water.
-
Place the transducer tip of the SonoSite Probe into the tube, and submerge to the 60cm mark
(minimum) as indicated on the probe shaft.
If using a soak tray (pictured, bottom-right):
-
Hang the ULT-PC-10 Conductivity Probe on the edge of the tray.
-
Fill the tray with cleaning solution, saline, or water until the Conductive Probe contacts are
sufficiently submerged (approximately 1 inch, or more).
-
Place the transducer tip of the SonoSite Probe into the tray, and submerge between 60cm and 90cm
as indicated on the probe shaft.
*If using an automated TEE Probe Disinfector see Appendices for details.
BC Biomedical, ULT-PC-30 or
ULT-PC-31 Submersible
Conductivity Probe
SonoSite Probe submerged between
60cm and 90cm as indicated on the
probe shaft.
Soak Tube Setup
Soak Tray Setup
BC Biomedical, ULT-PC-10
Conductivity Probe