BC Biomedical ULT-2000 series Quick Start Manual Download Page 4

 

Page 4 of 5 

ULT-2000 Series Quick Start Guide (SonoSite Ultrasound) 

 10/2019 

                                                                                           

 

If using a soak tube* (pictured, left): 

-

 

Insert the tip of the ULT-PC-30 Submersible Conductivity Probe to the bottom of the tube. 

-

 

Fill the tube to an appropriate level with cleaning solution, saline, or water. 

-

 

Place the transducer tip of the SonoSite Probe into the tube, and submerge to the 60cm mark 
(minimum) as indicated on the probe shaft. 

If using a soak tray (pictured, bottom-right): 

-

 

Hang the ULT-PC-10 Conductivity Probe on the edge of the tray. 

-

 

Fill the tray with cleaning solution, saline, or water until the Conductive Probe contacts are 
sufficiently submerged (approximately 1 inch, or more). 

-

 

Place the transducer tip of the SonoSite Probe into the tray, and submerge between 60cm and 90cm 
as indicated on the probe shaft. 

*If using an automated TEE Probe Disinfector see Appendices for details. 

 

 

BC Biomedical, ULT-PC-30 or 

ULT-PC-31 Submersible 

Conductivity Probe 

SonoSite Probe submerged between 

60cm and 90cm as indicated on the 

probe shaft. 

Soak Tube Setup 

Soak Tray Setup 

BC Biomedical, ULT-PC-10 

Conductivity Probe 