BC Biomedical PS-2010 User Manual Download Page 10

 

   

 
This section looks at the layout of a PS-2010 and gives descriptions of the elements that 

are present. 

  
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
The unit is controlled by 4 light touch keys.  They allow the user to select waveforms and 

control the power for the unit.  There is an audio click when a key is depressed. 

 

There  are  10  LEDs  to  provide  the  user  with  information  about  waveforms  that  are 

generated; 1 LED for Lead Test confirmation and 1 LED for Low Battery Indication. 

LAYOUT 

High Level 

Output (+) 

Lead Test 
Terminals 

4 Light Touch 
Keys for Power 
and Waveform 
Selection: 
•  On 
•  Off 
•  Up 
•  Down 

Universal Patient 
Lead Connectors: 
• V2/C2 
• V3/C3 
• V4/C4 
• V5/C5 
• V6/C6 
• RA/R 
• LA/L 
• RL/N  (-) 
• LL/F 
• V1/C1 
 

Lead Test 
Indicator LED 

Low Battery 
Indicator LED 

9V Battery 
Compartment 
     (Back) 
 
 

Battery 
Eliminator 
Receptacle (not 
shown) 

10 Waveforms with 
LED Indicators for 
Selection: 
 
ECG Selection: 
• 30, 60, 120, 240   
  BPM 
Performance Wave 
Selection: 
• Sine: 
 

10, 60, 100 Hz 

• Square: 
     0.125, 2.000 Hz 
• Triangle: 

2.000 Hz 

Summary of Contents for PS-2010

Page 1: ...ECG PATIENT SIMULATOR TEN LEAD PS 2010 USER MANUAL...

Page 2: ......

Page 3: ...7 NOTES 8 BC BIOMEDICAL PS 2010 TABLE OF CONTENTS CALIBRATION INTERVAL To ensure the accuracy of the PS 2010 Series BC Group International Inc recommends that it be calibrated at least once every 12...

Page 4: ...WARNING USE The PS 2010 is intended for testing only and should never be used in diagnostics treatment or any other capacity where it would come in contact with a patient WARNING USERS The PS 2010 Se...

Page 5: ...t the performance of the PS 2010 Allow the PS 2010 to acclimate to specified conditions for at least 30 minutes before attempting to operate it CAUTION LIQUIDS Do not submerge or spill liquids on the...

Page 6: ...iv...

Page 7: ...not directly connected to MAINS Under no circumstances should the terminals of the Analyzer be connected to any MAINS voltage NOTICE ABBREVIATIONS AHA American Heart Association ANSI American Nationa...

Page 8: ...E RIGHT TO MAKE CHANGES TO ITS PRODUCTS OR SPECIFICATIONS AT ANY TIME WITHOUT NOTICE IN ORDER TO IMPROVE THE DESIGN OR PERFORMANCE AND TO SUPPLY THE BEST POSSIBLE PRODUCT THE INFORMATION IN THIS MANUA...

Page 9: ...ECTORS ECG 30 60 120 AND 240 BPM SINE 10 60 AND 100 Hz SQUARE 0 125 AND 2 000 Hz TRIANGLE 2 000 Hz HIGH LEVEL OUTPUT 1 V p p AMPLITUDE ACCURACY 2 FREQUENCY ACCURACY 0 5 OF SETTING 9 VOLT BATTERY POWER...

Page 10: ...ted 1 LED for Lead Test confirmation and 1 LED for Low Battery Indication LAYOUT High Level Output Lead Test Terminals 4 Light Touch Keys for Power and Waveform Selection On Off Up Down Universal Pati...

Page 11: ...the High Level and RL N Patient Lead Connectors Waveform Selection There are two keys and 10 LEDs in the Waveform Selection Control Section The LEDs indicate which waveform is generated The keys seque...

Page 12: ...The unit will automatically turn off after 10 minutes of no key activity to conserve the battery To override this feature and keep the unit on continuously depress the key twice after the unit is on T...

Page 13: ...ttery Eliminator Input A 2 1 mm receptacle is provided for the optional 9 VDC Battery Eliminator power supply that may be used for continuous run applications It bypasses the internal battery when plu...

Page 14: ...E INTENDED THIS WARRANTY IS EFFECTIVE FOR TWELVE MONTHS FROM THE DATE OF SHIPMENT EXCLUSIONS THIS WARRANTY IS IN LIEU OF ANY OTHER WARRANTY EXPRESSED OR IMPLIED INCLUDING BUT NOT LIMITED TO ANY IMPLIE...

Page 15: ...IMPEDANCE LEAD TO LEAD 1000 LEAD TEST 1000 PHYSICAL ENVIRONMENTAL CONSTRUCTION ENCLOSURE ABS Plastic FACEPLATE Lexan Back printed SIZE 7 26 x 4 46 x 1 51 Inches 184 4 x 113 3 x 38 4 mm WEIGHT 1 Lbs 0...

Page 16: ...8 NOTES...

Page 17: ...9 NOTES...

Page 18: ...10 NOTES...

Page 19: ......

Page 20: ...INTERNATIONAL INC 3081 ELM POINT INDUSTRIAL DRIVE ST CHARLES MO 63301 USA 1 800 242 8428 1 314 638 3800 www bcgroupintl com sales bcgroupintl com PS 2010 User Manual 12 21 Rev 06 Copyright 2021 Made i...

Reviews: