BC Biomedical CQM-2000 Series User Manual Download Page 6

 

 

This manual covers the model CQM-2000 and CQM-2100 REM/ARM/CQM Test Module. 
The term CQM-2000 will be used to apply to both models. Sections specific to the CQM-
2100 model will use the term CQM-2100 exclusively. 
 
The  CQM-2000  is  used  to  test  the  Return  Electrode  Contact  Quality  Monitor  of  an 
Electrosurgical  Unit.    The  CQM-2000  provides  an  adjustable  resistance  from  0  to  210 
ohms in 1 ohm increments.  The CQM-2100 provides an adjustable resistance from 0 to 
1110 ohms in 1 ohm increments.  

The internal switches are ‘make-before-break’ type to 

ensure that the CQM circuit in the DUT does not see a momentary open condition.   

 
 

To Operate the Unit: 

  Connect the DUT Return electrodes to the CQM-2000 Test Jacks. 

  Set the Resistance using the knobs on the CQM-2000 Front Panel.   

 
 
Many ESUs measure a baseline resistance as the CQM connector is plugged in. 
 
 
The DUT should alarm based on the following conditions: 

  If the Resistance is too low, a CQM alarm should occur.   

  If the Resistance is too high, a CQM alarm should occur.   

  If the Resistance changes some percentage above or below the baseline 

 

resistance, a CQM alarm should occur.   

 
 
Baseline  Resistance  limits  and  percent  resistance  rise  limits  vary  based  on  ESU 
manufacturer.  Consult DUT manual for these parameters.   
 

 

BC BIOMEDICAL 

CQM-2000 Series 

REM/ARM/CQM TEST MODULE 

Summary of Contents for CQM-2000 Series

Page 1: ...REM ARM CQM TEST MODULE CQM 2000 Series USER MANUAL...

Page 2: ...TES 4 BC BIOMEDICAL CQM 2000 Series TABLE OF CONTENTS CALIBRATION INTERVAL To ensure the accuracy of the CQM 2000 BC Group International Inc recommends that it be calibrated at least once every 12 mon...

Page 3: ...t is connected when testing with the CQM 2000 Do not connect any leads from the patient directly to the CQM 2000 or DUT WARNING LIQUIDS Do not submerge or spill liquids on the CQM 2000 Do not operate...

Page 4: ...cloth A mild detergent can be used if desired CAUTION INSPECTION The CQM 2000 should be inspected before each use for obvious signs of abuse and wear The CQM 2000 should not be used and should be serv...

Page 5: ...THIS MANUAL HAS BEEN CAREFULLY CHECKED AND IS BELIEVED TO BE ACCURATE HOWEVER NO RESPONSIBILITY IS ASSUMED FOR INACCURACIES NOTICE CONTACT INFORMATION BC BIOMEDICAL BC GROUP INTERNATIONAL INC 3081 ELM...

Page 6: ...at the CQM circuit in the DUT does not see a momentary open condition To Operate the Unit Connect the DUT Return electrodes to the CQM 2000 Test Jacks Set the Resistance using the knobs on the CQM 200...

Page 7: ...OM THE DATE OF SHIPMENT EXCLUSIONS THIS WARRANTY IS IN LIEU OF ANY OTHER WARRANTY EXPRESSED OR IMPLIED INCLUDING BUT NOT LIMITED TO ANY IMPLIED WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR...

Page 8: ...fter subtraction of zero resistance CONNECTIONS Safety Jack 4 mm ENVIRONMENTAL AND PHYSICAL OPERATING RANGE 15 to 40 C 59 to 104 F STORAGE RANGE 20 to 65 C 4 to 149 F CONSTRUCTION Enclosure ABS Plasti...

Page 9: ...4 NOTES...

Page 10: ...NOTES...

Page 11: ...NOTES...

Page 12: ...RNATIONAL INC 3081 ELM POINT INDUSTRIAL DRIVE ST CHARLES MO 63301 USA 1 800 242 8428 1 314 638 3800 www bcgroupintl com sales bcgroupintl com CQM 2000 Series User Manual 01 20 Rev 05 Copyright 2020 Ma...

Reviews: