4 Check after Repair
4 - 2
Infusomat® compact
plus
P 1.0
EN
For internal use, only
PUMP RUN IN
Performing the RUN IN
1. To run-in the drive, insert an Intrafix Primeline
Do not close the pump door, yet.
2. Use the "Menu" button and the arrow keys to select the
"Settings" menu item and confirm with "OK".
3. Select "Service" and confirm the device code with "OK".
4. Confirm selection of "All Disposables Off" with "OK" (
5. Select "On" with "OK".
"Service active" is displayed.
6. Confirm " Do not use at Patient" with "OK" (
7. Close the pump door.
NOTICE
The "All Disposables On" function has to be activated before the
pump door is closed.
Fig. 4 - 1
Fig. 4 - 2