9. TROUBLE SHOOTING 

Problem

 

Probable cause

 

Action

 

   

    Fault code «E01» 

    +                       

blinks 

1.Motor is stuck 
2.Motor is damaged 
3.Blower  structure  failure 
caused by outer force 
4.Circuit failure 

Check  and  remove  physical 
failure  and  restart  the  system. 
Return  to  dealer  if  LCD  still 
shows E01 

    Fault code «E02» 

    +                 

blinks 

   

1.Motor is damaged 
2.Motor impeller rubs blower 
shell 
3.Circuit 

has 

excessive 

current. 

Check  and  remove  physical 
failure  and  restart  the  system. 
Return  to  dealer  if  LCD  still 
shows E02 

blinks 

    + blinks 

    + alarm sounds 

Low battery 

 

Charge the battery 

 

blinks 

    + blinks 

    + alarm sounds 

Filter blocked 
Tube blocked 

Remove obstruction, change the 
filter 
Clean tube   

 

alarm sounds 

 

Battery high temperature 

 

Stop working and rest 

No air flow, no alarm 

1. No power 
2.Battery contact damaged 

Charge the battery 
Check battery contact 

Battery run time is too short 

1.Battery is not fully charged 
2. Filter is blocked 
3. Battery is damaged 

Charge the battery 
Remove  obstruction,  change 
filter 
Change battery 

Air  supply  to  hood  smells   
unusual 

1. Filter broken   
2. Tube broken 
3. Hood damaged     

Leave current area immediately. 
1. Change filter   
2. Change tube 
3. Change hood 

Supply insufficient air to hood 

1. Breathing tube broken off 
2. Breathing tube broken   
3. Filter is blocked 

1. Check tube connection to hood 
and Respirator system unit 
2. change breathing tube 
3.Remove  obstruction,  change 
filter 

Summary of Contents for PeRSo 3.0

Page 1: ...PeRSo 3 0 User Manual Powered Air Purifying Respirator Please read these instructions carefully before using the equipment Incorrect use can result in risk to life and health...

Page 2: ...athing protection device It protects against particulate contamination All components used in the Respiratory System must be manufacturer approved parts and must be used in accordance with the instruc...

Page 3: ...evice is in the power off state a rapid build up of carbon dioxide and depletion of oxygen within the hood may occur The filters shall only be fitted to the blower unit and not directly to the helmet...

Page 4: ...parts must be replaced before use FIGURE 1 The package must include 1 The air fed hood 2 The respiratory protection system blower unit filter waist belt 3 Lithium ion battery 4 The tube and end fittin...

Page 5: ...filter cover is released 3 Remove the used filter by lifting it out from the filter cover 4 5 Remove the pre filter The pre filter and filter expected lifetime is 12 months When under intensive use c...

Page 6: ...e used for anything other than it was designed for Do not charge the battery in a potentially explosive area The charger must only be used indoors The charger regulates the charge automatically once t...

Page 7: ...sconnect the charger from the mains do not keep the charger plugged in to the mains if it s not in use 4 3 Installing the respirator system on the belt 1 Remove the belt s release buckle 2 Remove the...

Page 8: ...ist it clockwise to lock its position 2 Connect the other end of the tube to the headgear in the same way 1 Check that the respirator tube is firmly connected If the tube is broken replace it All comp...

Page 9: ...flow is sufficient if the marble reaches the minimum flow level O 1 The airflow must be tested before using 2 If the marble can t reach the minimum flow level don t use the system Change the filter o...

Page 10: ...show the PERSO 3 0 operating status FIGURE 3 FIGURE 3 Part 1 shows the current air flow rate Part 2 shows the level of the airflow Part 3 shows the filter condition Part 4 shows the battery Part 5 sh...

Page 11: ...is at level 3 230L min Press the ON button once again the air flow reverts to level 1 170L min 1 To turn off the blower unit long press the OFF button for more than 3 seconds 2 The Respirator system...

Page 12: ...P3 04 03T 001 Pre filter 4 P3 04 04T 001 Filter 5 P3 04 05T 001 Blower Unit 6 P3 05 01T 001 Waist Belt 7 P3 05 02T 001 Shoulder Harness 8 P3 03 01T 001 Standard Battery P3 03 01T 002 Heavy Duty Batte...

Page 13: ......

Page 14: ...external surfaces Don t use water The filter should be replaced together with the pre filter 8 2 Storage The Respirator system must be stored in a dry clean area in the temperature range of 10 to 55...

Page 15: ...d Tube blocked Remove obstruction change the filter Clean tube blinks alarm sounds Battery high temperature Stop working and rest No air flow no alarm 1 No power 2 Battery contact damaged Charge the b...

Page 16: ...ture 14 to 131 10 55 Battery Type Standard Battery Rechargeable Li ION 4400mAh Optional Heavy Duty Battery Rechargeable Li ION 6800mAh Expected Battery Operation Time Standard Battery Level 1 8h Level...

Page 17: ...lter type P Particle filter R Reusable type of particle filters and SL reflects the filter has been tested against particles of liquid and solid matter Warning sound indication This PERSO 3 0 PAPR has...

Page 18: ...om the date of purchase The company undertakes to exchange or repair without charge any part found to be defective within this period alternatively and at its discretion The Company may replace This g...

Reviews: