Carefully read all instructions prior to use.  Observe all contraindications, warnings and 
precautions noted in these instructions.  Failure to do so may result in patient complications.

 

 
Caution: Federal (U.S.A.) law restricts this device to sale by or on the order of a physician. 
 

I. 

DEVICE DESCRIPTION  

The RFP-100A Footswitch (Footswitch) is used with the RFP-100A Baylis Medical Company 
Radiofrequency Puncture Generator (RFP-100A Generator) for RF energy delivery. The 
Footswitch is provided with redundant micro switches and is used as an alternative to the RF 
On/Off button on the RFP-100A Generator for RF energy delivery. 

II. 

INDICATIONS FOR USE  

The Footswitch is to be used with the RFP-100A Generator. 

III. 

CONTRAINDICATIONS   

The  Footswitch  is  not  recommended  for  uses  other  than  the  indicated  use.  The  RFP-100A 
Generator Instructions for Use should be consulted for any other contraindications that may be 
associated with use of the RFP-100A Generator. 

IV. 

WARNINGS  

 

Only  physicians  with  a  thorough  understanding  of  radiofrequency  powered  puncture 
procedures should use this device. 

 

Do not alter this device in any way. 

 

The Footswitch must be used with the RFP-100A Generator only.  Attempts to use it with 
other  radiofrequency  puncture  generators  can  result  in  serious  electrical  injury  of  the 
patient and/or operator. 

V. 

PRECAUTIONS  

 

Do  not  attempt  to  use  the  Footswitch  before  thoroughly  reading  the  accompanying 
Instructions for Use and RFP-100A Generator Instructions for Use manual. 

 

Visually inspect  the  Footswitch  prior  to  use.   Do  not  use  the  Footswitch if  there is  any 
damage.  

 

Regularly inspect and maintain the Footswitch according to section VII. and X.   

VI. 

ADVERSE EVENTS  

The RFP-100A Generator Instructions for Use should be consulted for any adverse 
events. 

VII. 

INSPECTION PRIOR TO USE 

Prior to use all the individual components including the RFP-100A Generator and the Footswitch 
should  be  carefully  examined  for  damage  or  defects,  as  should  all  equipment  used  in  the 
procedure.  Do not use defective equipment. 

VIII. 

DIRECTIONS FOR USE  

The Footswitch serves as an actuator for the connected device. When connected to the RFP-
100A Generator, the Footswitch can be used as an alternative to the RF ON/OFF button on the 
RFP-100A Generator to activate or terminate RF energy delivery.  
For Footswitch procedure and connections to the RFP-100A Generator please refer to the RFP-
100A Generator Instructions for Use manual. 

IX. 

CLEANING INSTRUCTIONS 

The outer surface of the Footswitch may be cleaned with a mild soapy solution. DO NOT 
immerse the Footswitch in any liquid.  Avoid caustic or abrasive cleaners.  If disinfecting is 
required, 70% isopropyl alcohol or 5% solution of household bleach may be used to clean the 
outer surfaces. The Footswitch cannot be sterilized.  

X. 

MAINTENANCE 

Regular checks or maintenance is recommended according to ambient conditions and 
application frequency:  
Check actuating elements for mechanical function. To perform this test, disconnect the 
footswitch from RFP-100A Generator. Press on the footswitch and release force. Confirm that 
the footswitch returns to its rest position (released state). 
Check housing and connecting cable for damage and harmful dirt. 

XI. 

CUSTOMER SERVICE AND PRODUCT RETURN INFORMATION  

If  you  have  any  problems  with  or  questions  about  Baylis  Medical  Equipment  contact  our 
technical support personnel.  

NOTES: 

1. 

In order to return products you must have a return authorization number before shipping 
the products back to Baylis Medical Company. 

2. 

Baylis  Medical  will  not  accept  any  piece  of  used  equipment  without  a  sterilization 
certificate. . 

XII. 

LABELING AND SYMBOLS 

 

Manufacturer 

 

Consult 
Instruction
s for Use   

 

EU Authorized 
Representative 

            

           

Model 
number  

 

Caution 

  

 

Serial 
Number 

 

Only for EU member states: 
Use of this symbol indicates that the product must be 
disposed of in a way that complies with local and national 
regulations. For questions regarding recycling of this 
device please contact your distributor 

 

Federal (U.S.A.) law restricts this device to sale by or on 
the order of a physician. 

XIII. 

Technical Data 

Standards 

IEC 60601-1 
IEC 60601-1-2 
IEC 60601-2-2 
IEC 60601-2-22 
IEC 60529 

Enclosure 

Unbreakable, self extinguishing thermoplast. 

Contact 

Reed 

Contact Material 

Rhodium 

Protection Class 

IP X8 as per IEC 60529 

Electrical Connection 

3 m shielded cable (3xAWG 24C UL, GY) with LEMO-
Connector 4 

–pole. 

Electrical Characteristics: 

 

Baylis Medical Company Inc. 
5959 Trans-Canada Highway 
Montreal, Quebec, Canada, H4T 1A1 
Tel: (514) 488-9801/ (800) 850-9801 Fax: (514) 
488-7209 
www.baylismedical.com 

 

 

 

 

 

EU Authorized Representative: 

Quality First International 
20 Eversley Road, Bexhill-on-Sea, East Sussex 
TN40 1HE, United Kingdom 
Tel: +44-(20)-8-522-1937 Fax: +44-(20)-8-522-1937 

Instructions for Use 

 

RFP-100A Footswitch 

 

 

Summary of Contents for RFA-FS

Page 1: ...he Footswitch can be used as an alternative to the RF ON OFF button on the RFP 100A Generator to activate or terminate RF energy delivery For Footswitch procedure and connections to the RFP 100A Gener...

Page 2: ...bien les proc dures de ponction par radiofr quence doivent utiliser ce dispositif Ne modifiez pas ce dispositif de quelque mani re que ce soit L interrupteur au pied doit tre utilis exclusivement ave...

Page 3: ...100A este pedal se puede utilizar para conectar o desconectar el suministro de energ a de RF como alternativa al bot n de activaci n desactivaci n de RF que se encuentra en el generador RFP 100A Para...

Page 4: ...canso estado libertado Verificar a estrutura exterior e o cabo de liga o para detetar se existem danos e sujidade prejudicial XI ATENDIMENTO AO CLIENTE E INFORMA ES SOBRE DEVOLU O DO PRODUTO Se tiver...

Page 5: ...1 IEC 60601 1 2 IEC 60601 2 2 IEC 60601 2 22 IEC 60529 Involucro Termoplastica infrangibile e autoestinguente Contatto Canna Materiale di contatto Rodio Classe di protezione IP X8 per IEC 60529 Colleg...

Page 6: ...aden Deze beperkte garantie is alleen van toepassing op nieuwe originele producten zoals geleverd vanaf de fabriek en die gebruikt zijn voor normale en beoogde doeleinden BMC s beperkte garantie is NI...

Page 7: ...lative Luftfeuchte 10 100 10 100 Luftdruck 500 hPa 1060 hPa 700 hPa 1060 hPa XIV BEGRENZTE GARANTIE Einwegartikel und Zubeh r Baylis Medical Company Inc BMC gew hrleistet eine Garantie f r seine Einwe...

Page 8: ...RLITTEN WERDEN HAFTBAR ZU MACHEN IST DIE HAFTBARKEIT VON BMC IST BEGRENZT AUF DIE KOSTEN DES K UFERS F R DEN KAUF DER SPEZIELLEN WAREN DIE VON BMC AN DEN K UFER VERKAUFT WURDEN UND DIE DEN GRUND F R D...

Reviews: