15
You must place the test
strip while the message
PLACE STRIP ON TABLE is
on the screen. It will appear
for 10 seconds.
5
The table is automatically
pulled into the instrument for
reading. Results are available in
one minute. Be sure not to move
or bump the table.
3
Blot the strip to remove
excess urine by touching the
edge to a paper towel. Do not
drag the strip across the towel;
touch the edge only.
4
Place the Reagent Strip,
with the test pads facing up,
into the middle trough of the test
strip table. Slide the strip along
the table until it touches the end
of the trough.