Injection monitoring is performed in the injector head to detect over rate or over volume con-
ditions due to system faults. The delivered volume is also monitored against the total pro-
grammed volume for the injection.
•
When a fault condition, hold or stop is detected, the injection will stop within 5 mL.
•
Once the system has disarmed, a tone sounds and a stall message displays on the display
screen.
17.7 System Fluid Performance
The performance of the injection system has been tested with disposables supplied by Bayer with a
pressure rating greater than 325 psi (2241 kPa) and the catheters in Table 17 - 2.
17.7.1 Factors Affecting Flow Rates
An injector's ability to generate pressure is only one factor affecting maximum flow rates. Other factors
include:
•
Catheter diameter
•
Viscosity (thickness) of the fluid
•
Tube length
•
Number of tubing sections
•
Tube diameter
•
Temperature of the fluid, tube, and syringes during the injection
•
Maximum pressure setting on the injector
If problems are experienced achieving the desired flow rate, please contact a Clinical representative for
suggestions that may increase it.
17.7.2 Maximum Flow Rate Performance
Maximum flow rates were achieved using the following catheters, with inner diameters (ID) and
lengths as noted. Contrast was heated to 37° C ± 1° C. System programmable pressure limit of 325
psi (2241 kPa) was used for testing.
NOTE:
Individual results may vary. Maximum flow rate performance results were met using the
components and conditions described in this section. Different components and/or conditions
may lead to different results.
Table 17 - 2: Steady State Maximum Flow Rate Using 60-inch LPCT from Bayer
Catheter
Fluid
18 Gauge
ID:
0.876 mm – 0.978 mm
Length:
2.95 cm (1.16 in.)
20 Gauge
ID: 0.800 mm
Length: 2.54 cm (1.00 in.)
22 Gauge
ID: 0.546 mm – 0.648 mm
Length: 2.54 cm (1.00 in.)
24 Gauge
ID: 0.445 mm – 0.521 mm
Length: 1.91 cm (0.75 in.)
Ultravist 300
8.5 mL/s
7.0 mL/s
6.0 mL/s
4.0 mL/s
Ultravist 370
8.5 mL/s
7.0 mL/s
5.0 mL/s
3.0 mL/s
0.9% Saline
8.5 mL/s
7.0 mL/s
6.0 mL/s
4.0 mL/s
Summary of Contents for Certegra
Page 2: ......
Page 4: ...MEDRAD Stellant CT Injection System with Certegra Workstation ii ...
Page 18: ...MEDRAD Stellant CT Injection System with Certegra Workstation 2 10 ...
Page 34: ...MEDRAD Stellant CT Injection System with Certegra Workstation 6 26 ...
Page 36: ...MEDRAD Stellant CT Injection System with Certegra Workstation 7 28 ...
Page 54: ...MEDRAD Stellant CT Injection System with Certegra Workstation 10 46 ...
Page 58: ...MEDRAD Stellant CT Injection System with Certegra Workstation 11 50 ...
Page 60: ...MEDRAD Stellant CT Injection System with Certegra Workstation 12 52 ...
Page 70: ...MEDRAD Stellant CT Injection System with Certegra Workstation 13 62 ...
Page 86: ...MEDRAD Stellant CT Injection System with Certegra Workstation 16 78 ...
Page 103: ......