Preparing for Injection
•
•
9.1 Control Room Preparation
9.1.1 Informatics Panel
Refer to Certegra
®
Applications and Workstation Accessories manual.
9.1.2 Prepare Injection Protocol
1.
Set protocol.
a.
To recall a protocol:
Figure 9 - 1: Recall a Protocol
i.
On the Home screen, select Protocol Manager.
(1)
ii.
Select View All to see all saved protocols.
(2)
iii.
Select applicable protocol from list.
(3)
iv.
Review all protocol values.
(4)
v.
Select OK to use or edit the protocol.
(5)
vi.
Refer to
for further information.
b.
To create or edit a protocol:
Vessel Hazard - Serious patient injury or death may result.
•
Ensure that the programmed flow rate meets facility guidelines.
Minor or moderate patient injury may result.
•
Confirm that the programmed protocol is correct prior to selecting Lock on the display.
WARNING
CAUTION
1
2
3
4
5
Summary of Contents for Certegra
Page 2: ......
Page 4: ...MEDRAD Stellant CT Injection System with Certegra Workstation ii ...
Page 18: ...MEDRAD Stellant CT Injection System with Certegra Workstation 2 10 ...
Page 34: ...MEDRAD Stellant CT Injection System with Certegra Workstation 6 26 ...
Page 36: ...MEDRAD Stellant CT Injection System with Certegra Workstation 7 28 ...
Page 54: ...MEDRAD Stellant CT Injection System with Certegra Workstation 10 46 ...
Page 58: ...MEDRAD Stellant CT Injection System with Certegra Workstation 11 50 ...
Page 60: ...MEDRAD Stellant CT Injection System with Certegra Workstation 12 52 ...
Page 70: ...MEDRAD Stellant CT Injection System with Certegra Workstation 13 62 ...
Page 86: ...MEDRAD Stellant CT Injection System with Certegra Workstation 16 78 ...
Page 103: ......