background image

What Y

ou See

What It Means

What Y

ou Should Do

Low Battery symbol 

fl 

ashes if battery power 

is low

.

Replace the battery (CR2032) 

within 20 r

eadings or one week.

Battery life has ended.

Replace the battery (CR2032).

Temperatur

e is outside 

the operating range of 

the meter (too hot or 

too cold).

Allow the meter to r

each r

oom 

temperatur

e.

Meter does not 

recognize the test disc. 

Remove the test disc and inspect 

the meter for for

eign material. 

Rotate the disc and r

einstall it. If 

err

or r

epeats, try inserting a new 

disc. If err

or continues, contact 

Bayer Customer Service.

Test strip was r

emoved 

during the test, or the 

release button was 

pr

essed during the test.

Take car

e to not pr

ess the r

elease 

button or r

emove the test strip 

during the test. Release the used 

test strip and r

epeat the test (see 

page 1

0). If the err

or continues, 

contact Bayer Customer Service.

(with continuous 

beep)

The meter was open 

when the meter handle 

was pulled. 

Close the meter and push the 

handle in.

What Y

ou See

What It Means

What Y

ou Should Do

1. 

You may have 

 

applied blood too 

 

early to the test strip. 

2. 

Ther

e may be 

 

moistur

e in the test 

 strip 

slot.

1. 

Repeat the test and wait until 

 

you see the “apply blood” 

 

pr

ompts befor

e you touch the 

 

strip to the blood dr

op.

2. 

Wick the moistur

e away (see 

 

page 36), then r

epeat the test.

Ther

e was a pr

oblem 

completely fi

 lling 

the 

test strip; accurate test 

result is not possible. 

Release the used test strip and 

repeat the test with a new strip 

(see page 1

0). 

1.  

Meter has detected 

an err

or; accurate 

result is not possible. 

2.  

The 1

0-test disc may 

be past its expiration 

date or may have 

been damaged due 

to intense heat or 

moistur

e exposur

e. 

1.  

Release the used test strip and 

repeat the test with a new strip 

(see page 1

0).

2.  

Replace the 1

0-test disc, using a 

new box if necessary

.

Your blood glucose level

 

is very high or the meter

 

has detected an err

or;

 

accurate test r

esult is

 

not possible.

Turn the meter 

of

f,

 then 

on

 using 

. Retest. If the err

or appears 

again, call your physician or 

healthcar

e pr

ofessional. Y

our 

blood glucose may be above 

83.3 mmol/L. If the pr

oblem is 

with the meter

, contact Bayer 

Customer Service.

Bayer Customer Ser

vice: (852) 8

100 6386

44

45

www

.bayerdiabetes.com

82076779_BRZ2_UG_EN_FpBp_v0.indd   44-4582076779_BRZ2_UG_EN_FpBp_v0.indd   44-45

8/8/11   12:45:48 PM8/8/11   12:45:48 PM

Summary of Contents for Bai An Jie 2

Page 1: ...PEC V 0 ENGLISH HONG KONG PROCESS CYAN PROCESS MAGENTA PROCESS YELLOW PROCESS BLACK COLORS Avenir Berthold Akzidenz DFPGYaYi DFPGBiaoSong Futiger Futura Helvetica Times FONTS August 8 2011 Bayer HealthCare Breeze 2 Rebranding UG English 907186 82076779 Rev 07 11 N A 6 0 H x 4 5 W PUSG0494 REV F 82076779_BRZ2_UG_EN_FpBp_v0 indd ofc1 82076779_BRZ2_UG_EN_FpBp_v0 indd ofc1 8 8 11 12 45 16 PM 8 8 11 12...

Page 2: ...R UNNING A B LOOD G LUCOSE T EST 10 MARKING A DELETED BLOOD TEST 18 UNDERSTANDING YOUR TEST RESULTS 19 ALTERNATIVE SITE TESTING 20 R UNNING A C ONTROL T EST 22 MARKING A CONTROL TEST 24 B EYOND E VERYDAY U SE R ECALLING T EST R ESULTS 26 VIEWING YOUR AVERAGES 27 CLEARING STORED TEST RESULTS 28 S ETUP F UNCTION 29 SETTING THE TIME 29 SETTING THE DATE 30 SETTING THE BUZZER LEVEL 32 SETTING THE TIME ...

Page 3: ...is divided into color coded sections to help you easily fi nd the information you need See the Table of Contents for a complete list of topics CAUTION Before using any product to test your blood glucose sugar read all instructions carefully and practice the test T ECHNICAL I NFORMATION S PECIFICATIONS 40 P ERFORMANCE E VALUATIONS 41 S OLVING P ROBLEMS S OLVING P ROBLEMS 42 T ROUBLESHOOTING AND E RR...

Page 4: ...Setup mode Button Door Open Open Latch Press in and pull up on the base to open the meter Back Base of Meter NOTE Simply place the easy reference sticker on the button door as shown See the sheet of stickers provided in your meter package and select the language of your choice On Off Button Used to manually turn the meter on or off check the number of strips and check the date time Memory Button U...

Page 5: ...ble shooting and Error Codes on page 43 Here is what each display element means The units of measure mg dL or mmol L for your blood glucose result Where the time date and AM PM are displayed also the type of average Here is what each display element means Additional results and options are available by pressing the and buttons Battery is low or dead Apply blood to the test strip The number of test...

Page 6: ... on the disc T ESTING M ATERIALS щଥ Bai An Jie 10 Test Disc Disc contains ten test strips Not included with the kit щଥ Bai An Jie Control Solution Not included with the kit M ICROLET 2 Lancing Device and M ICROLET Lancets Included with the kit WARNING The M ICROLET 2 lancing device is intended for self testing by a single patient It must not be used on more than one person due to the risk of infec...

Page 7: ... notches in the disc with the tabs on the meter Close Meter 4 Close the meter and snap it shut Keep the meter fl at when closing it Counting the Number of Strips Left Your щଥ Bai An Jie meter will automatically count down the number of test strips you have left The number is shown with your test result To see the count when the meter is off press the button NOTE This count is reset to 10 every time...

Page 8: ...ly into the lancing device until it comes to a full stop This will cock the device You can also pull and release the cocking handle Twist off the lancet cap Be sure to save it for disposing of the used lancet 4 Replace the gray endcap prior to getting your blood drop Push Out Test Strip 5 Your щଥ Bai An Jie meter uses a special easy to use Pull and Push design Here s how it works Hold the meter co...

Page 9: ... re insert a test strip into the meter NOTE After the meter handle is pushed in the release button is UP a test strip is pushed out and a series of apply blood prompts appears The meter is now ready to run a blood glucose test 9 Your test strip is like a straw it sips in a sample of blood Move the entire front edge of the test strip until it touches the edge of the blood drop The blood is automati...

Page 10: ... may not release properly 12 Press the button to turn the meter off NOTE If you forget to turn your meter off it will turn off automatically after three minutes This will save battery power Your test result then appears on the screen Helpful Hint The number displayed in the upper left cor ner is the number of test strips remaining in your 10 test disc IMPORTANT If you want to mark this result as a...

Page 11: ...liter Test results in mg dL will never include a decimal point Example or If your test result is displayed in the wrong unit of measure mg dL contact your authorized Bayer representative or call Customer Service see the bottom of the page Remove the used lancet 1 In one hand hold the lanc ing device with your thumb on the grip indent with your other hand hold the gray endcap dial as shown and gent...

Page 12: ... vomiting very dry mouth Low blood glucose hypoglycemia shakiness sweating fast heartbeat blurred vision strange behavior confusion passing out and seizure If you are experiencing any of these symptoms test your blood glucose If your result is below 2 8 mmol L or above 13 9 mmol L or is displayed as LO or HI contact your doctor immediately 1 American Diabetes Association Standards of Medical Care ...

Page 13: ...n steady pressure until the blood drop reaches the sug gested size 1 0 μL Lift the device away from the skin without smearing the blood Do not test the blood sample if you get Smeared blood Clotted blood Runny blood Clear fl uid mixed with the blood Bayer Customer Service 852 8100 6386 20 21 www bayerdiabetes com Unusual Test Results If your blood glucose result doesn t match the way you feel follo...

Page 14: ...a series of apply blood prompts appears The meter is now ready to run a control test 2 Squeeze a small drop of щଥ Bai An Jie control solution onto a nonabsorbent surface such as a clean piece of wax paper Do not apply control solution to the test strip directly from the bottle NOTE The control solution contains a red dye and may stain Test 3 Move the entire front edge of the test strip until it to...

Page 15: ...des it from being included in your test averages You must mark the result immediately after the test is performed while the result is still displayed on the screen To do this follow these steps 5 Open the button door Press the button to mark as a control The symbol indicates a control result Press to accept 6 To release the used test strip hold the meter with the test strip pointing down over a wa...

Page 16: ...t Viewing Your Averages Your щଥ Bai An Jie meter calculates several differ ent averages of your blood glucose results You can view averages for 1 day the previous 24 hours 7 days 14 days and 30 days To see the averages simply follow these steps 1 Open the button door and press the button The meter will show you your last result 2 Press the button again to see your blood glucose average for the las...

Page 17: ...e all three dashes have been lit S ETUP F UNCTION Helpful Hint You can press the button at any time during Setup to turn the meter off Just remember to press the button fi rst to store your changes Setting the Time 1 Open the button door see page 3 2 Press 3 Press or as needed to change the hour When the correct hour is being displayed press to accept it and move on 4 Press or to change the minutes...

Page 18: ...head to step 4 1 Open the button door 2 Press 3 Press until you see the day display fl ashing 4 Press or to change the day Press to accept and move on 5 Press or to change the month Press to accept and move on 6 Press or to change the year Press to accept and move on 82076779_BRZ2_UG_EN_FpBp_v0 indd 30 31 82076779_BRZ2_UG_EN_FpBp_v0 indd 30 31 8 8 11 12 45 43 PM 8 8 11 12 45 43 PM ...

Page 19: ...l 4 Press or to change the buzzer level Press to accept and move on Setting the Time Format 12 or 24 hour Clock Your meter can display the time format as either 12 hour or 24 hour You can change the time format as follows NOTE If you have just set the buzzer level skip ahead to step 4 1 Open the button door 2 Press 3 Press until you see the time format display 4 Press or to change the time format ...

Page 20: ...ave just set the time format skip ahead to step 4 1 Open the button door 2 Press 3 Press until you see the date format display 4 Press or to change the date format Press to accept T RANSFERRING R ESULTS TO A C OMPUTER You can transfer test results from the щଥ Bai An Jie meter to a computer where they can be summarized in a report with graphs and tables To make use of this feature you need Bayer s ...

Page 21: ...approximately 20 readings or one week R EPLACING THE B ATTERY Your щଥ Bai An Jie meter uses one 3 volt lithium battery CR2032 A battery will already be installed in the meter when you buy it A fl ashing battery on the screen means that it is time to put a new battery in the meter You will not erase test results in the meter s memory when you change the battery You may need to reset the time WARNING...

Page 22: ... rst in order to remove the battery holder DO NOT PRY the battery holder out 3 Remove the old battery by pushing it out from the back side with your thumb 4 Snap 1 new battery CR2032 fi rmly in the holder with the side up 5 Gently slide the battery holder back into the battery compartment Make sure the battery holder is pushed all the way in 6 Close the meter 7 Check the meter s time press the butt...

Page 23: ...curacy of a blood glucose system shows how closely that system s results compare to laboratory results One hundred ninety seven 197 subjects with diabetes punctured their fi ngers and tested their own capillary blood using the щଥ Bai An Jie meter system Eight healthcare professionals then performed fi nger punctures on the subjects and tested the capillary blood using the meter system Capillary bloo...

Page 24: ... 6 mmol L Repeat the test and make sure the test strip is completely filled If LO appears again contact your physician or healthcare professional Test result is above 33 3 mmol L Wash your hands and the test site Repeat the test If HI appears again contact your physician or healthcare professional You have not pushed the meter handle all the way in Push the meter handle in until it clicks You must...

Page 25: ...rly to the test strip 2 There may be moisture in the test strip slot 1 Repeat the test and wait until you see the apply blood prompts before you touch the strip to the blood drop 2 Wick the moisture away see page 36 then repeat the test There was a problem completely fi lling the test strip accurate test result is not possible Release the used test strip and repeat the test with a new strip see pag...

Page 26: ...e page 36 and remove debris if present Retest 3 Contact Bayer Customer Service What You See What It Means What You Should Do Control Test result is out of range too high or too low Control solution is past its expiration date or is past the 6 month open use life date 10 test disc is past its expiration date Deteriorated test strip due to heat or exposure to moisture Control solution is not at room...

Page 27: ...is printed on the test strip disc and the test strip disc box Lot number ___________ Expiration date ____________ R EPLACEMENT I TEMS Replacement Parts Part Number Item 40030030 1 lithium battery CR2032 also available at most stores 82076779 щଥ Bai An Jie User Guide 82076787 щଥ Bai An Jie Quick Reference Guide You can obtain these parts by calling 852 8100 6386 or by writing Bayer HealthCare Limit...

Page 28: ...t strip discs manufactured or recommended by Bayer 3 Bayer reserves the right to make changes in design of this instrument without obligation to incorporate such changes into previously manufactured instruments 4 Bayer has no knowledge of the performance of the щଥ Bai An Jie meter when used with any test strip discs other than щଥ Bai An Jie test strip discs and therefore makes no warranty of the p...

Page 29: ... Harbour Road Wanchai Hong Kong Bayer the Bayer Cross щଥ Bai An Jie MICROLET and the No Coding logo are registered trademarks of Bayer No Coding AUTODISC and VACULANCE are trademarks of Bayer 82076779_BRZ2_UG_EN_FpBp_v0 indd 52 82076779 BRZ2 UG EN FpBp v0 indd 52 8 8 11 12 45 50 PM 8 8 11 12 45 50 PM ...