13
Note: The release button is UP, a test strip is pushed out
and a flashing blood drop appears on the screen. The
meter is now ready to run a control test.
2. Squeeze a small drop of control solution onto a
nonabsorbent surface (such as a clean piece of wax paper).
Do not apply control solution to the test strip directly from
the bottle.
Note: The control solution contains a red dye and
may stain.
Test
3. Move the entire front
edge of the test strip to
touch the edge of the
drop of control solution.
The control solution is
automatically pulled
into the test strip.
Important: Hold the test
strip in place until the
meter beeps, and three
dashes appear on the
screen, then move it
away. By doing so, you
will make sure the test
strip is completely filled.