44
This may affect the way
you see your results.
If there is a difference in
your meter’s display and
the one shown on page v,
contact your Bayer
Customer Service
Department (see pg. 45).
Some display
segments do not
appear to be
working when
the meter
handle is all the
way out.
What You See
What It Means
What You Do
• Control solution is past
expiration date or is past
the 6-month open use-
life date.
• 10-test disc is past
expiration date.
• Deteriorated test strip
due to heat or exposure
to moisture.
• Control solution not at
room temperature.
• Not enough control
solution drawn into test
strip.
Run another control test.
If still out of range, retest
with a new 10-test disc and
control solution with an
acceptable expiration date
and open use-life date. If
still out of range, contact
your Bayer Customer
Service Department
(see pg. 45).
Control Test
result is out
of range (too
high or too
low).
Your meter may be dirty.
Clean meter (see pg. 27). If
you continue to experience
difficulty, contact your
Bayer Customer Service
Department (see pg. 45).
Meter shows
dashes before
blood is applied.
Invalid test result.
Discard strip and retest.