EN
USER’S MANUAL
BR16
62
After introducing the parameters LF2, the indicator will show the last configuration saved. All the
steps to follow must be completed, if they are not done the indicator will continue with the previous
configuration.
Proceed with the calibration of weight after LF2.
LF 2
•
Press the key
to start and display the internal value. Press
to exit the menu and the scale
will automatically restart.
262144
•
Press the key
To continue with the configuration. (press the key
to exit from the configu-
ration and go back to the menu LF2).
100001
•U
se the keys
,
and
,
0 ~ 9
and then press the key
to choose the UNITS
OF WEIGHT
.
(press the key
to exit from the configuration and go back to the menu LF2).
000000kg
•U
se the keys
,
and
,
0 ~ 9
and then press the key
to choose the MAXI-
MUM WEIGHT. press the key
to exit from the configuration and go back to the menu LF2).
d 0.0kg
•U
se the keys
,
and Then
To move the decimal point ( press the key
To exit
from the configuration and go back to the menu LF2).
div 01
•Use
and then
to change the step.
(Press the key
To exit from the configuration and
go back to the menu LF2).
LF2
•Use
,
and then
To continue with the other configuration or press the key
to
exit from the menu and the scale will begin again automatically).
SPEED OF THE AD CONVERTER
LF 4
SPEED 1
standard speed 15Hz.
SPEED 2
high speed 30Hz.
SPEED 3
low speed 7.5Hz
*This function stays blocked when UF-5 is in mode HOLD 1.
*The value of factory is 1
DISPLAY
DESCRIPTION AND SEQUENCE OF USE
LF4
•P
ress the key
to start or the key
to exit from the menu and the scale is going to begin
again automatically.
Summary of Contents for BR16
Page 65: ...65 BR16...