07-19-B1-729
FLO-GARD 6301 DUAL CHANNEL VOLUMETRIC INFUSION PUMP SERVICE MANUAL
7-19
Troubleshooting
Troubleshooting Chart
Troubleshooting Chart
Table 7-4 identifies specific failures by symptoms rather than Failure
Identification Codes.
Table 7-4
Troubleshooting Chart
Symptom
Checks
Corrective Action
Power Issues
One or both pumps cannot be turned
on.
Check backup battery voltage in
accordance with “Automatic Test
Mode 1: Calibration Mode 1” on page
9-14.
Check front panel cables for short
circuits and proper connection.
Check CN601 and CN602 for proper
connection and continuity.
Sometimes this condition is followed
by a F49 and/or F94 code. Charge or
replace the battery to clear the
condition. Repair if necessary.
One or both pumps cannot be turned
off.
Check backup battery voltage in
accordance with “Automatic Test
Mode 1: Calibration Mode 1” on page
9-14.
Check front panel cables for short
circuits.
Check CN601 and CN602 for proper
connection and continuity.
Check if all 3 LEDs are on and the
constant alarm is on.
Sometimes this condition is followed
by a F49 and/or F94 code. Recharge
or replace the battery to clear the
condition. The battery compartment
may need to be opened and the battery
momentarily disconnected to turn off
the pump.
Disconnect then reconnect lithium
battery.
Replace front panel film.
Replace CPU board.
Replace Sensor board.
Pump does not operate long on battery
power.
Verify condition and capacity of
battery.
Verify continuity of the battery
harness.
Charge the battery uninterrupted for at
least 16 hours.
Replace the battery.
Repair or replace the battery harness.
Battery does not provide rated backup
power when AC power is not
available.
Perform “Battery Check” on page
Recharge or replace battery as
required.
Pressing a
START
key causes a
BATTERY LOW alarm.
Power the pump off and on again.
If the problem persists, recharge the
battery.
Repair as required.
If the problem persists, replace the
battery.
Summary of Contents for Flo-Gard 6301
Page 8: ...vi FLO GARD 6301 DUAL CHANNEL VOLUMETRIC INFUSION PUMP SERVICE MANUAL 07 19 B1 729 ...
Page 12: ...x FLO GARD 6301 DUAL CHANNEL VOLUMETRIC INFUSION PUMP SERVICE MANUAL 07 19 B1 729 ...
Page 244: ...A A 6 FLO GARD 6301 DUAL CHANNEL VOLUMETRIC INFUSION PUMP SERVICE MANUAL 07 19 B1 729 ...
Page 251: ...07 19 B1 729 ...