10
FSR Check
10-8
FLO-GARD 6301 DUAL CHANNEL VOLUMETRIC INFUSION PUMP SERVICE MANUAL
07-19-B1-729
3.
Open the pump door and slightly press the safety clamp to position the
air bubble below the pump fingers and above the air sensor.
4.
Close the door and start the pump. When the intact air bubble reaches the
air sensor the following should occur when an alarm is expected:
•
AIR is displayed in the Message Display.
•
The red ALARM LED flashes.
•
The audible alarm is activated.
•
The pump stops.
Q
If the pump does not alarm as required, verify that the bubble did not
break up before reaching the sensor by observing the bubble
downstream from the sensor. Either repeat the test or recheck and
adjust the sensor calibration.
Q
If the pump alarms with the 25 or 50 microliter bubbles, recheck and
adjust the air sensor calibration.
5.
Test both MIN and NORM options and all four air bubble volumes.
Upon completion, return the pump to the correct air sensor
configuration.
6.
Repeat this procedure for the other pump.
FSR Check
1.
Note the left and right FSR values displayed while the pump is on.
2.
Note the left and right FSR values displayed while the pump is
unplugged.
3.
Note the difference between the two values from steps 1 and 2.
4.
Wait for one minute and plug the pump back into AC power.
5.
Note the left and right FSR values displayed when the pump is plugged
in:
Q
If the values are within 10 counts in the readings, proceed to the next
step.
Q
If there is a jump of more than 10 counts in the readings:
Summary of Contents for Flo-Gard 6301
Page 8: ...vi FLO GARD 6301 DUAL CHANNEL VOLUMETRIC INFUSION PUMP SERVICE MANUAL 07 19 B1 729 ...
Page 12: ...x FLO GARD 6301 DUAL CHANNEL VOLUMETRIC INFUSION PUMP SERVICE MANUAL 07 19 B1 729 ...
Page 244: ...A A 6 FLO GARD 6301 DUAL CHANNEL VOLUMETRIC INFUSION PUMP SERVICE MANUAL 07 19 B1 729 ...
Page 251: ...07 19 B1 729 ...