Operator Manual for the Baxter ExactaMix 2400 Compounder
Page 18
53007410 Rev. C
ORDERING SUPPLIES
Order supplies through your normal channels.
Baxter Customer Service at +1.800.567.2292.
WARNING
Use only sterile inlets, bags and valve sets validated by Baxter.
INLETS
The following inlets are available worldwide.
Product
Order
Number
Quantity /
Case
For use with:
Standard
Priming
Volume
Minimum
Priming
Volume
Non-vented High-
Volume Inlet
H938
173
25
Large-volume, vented or collapsible
containers (such as bags of dextrose
and water)
50–60
mL
25–30 mL
Vented High-
Volume Inlet
H938
174
25
Large-volume, non-vented
containers that require a spike to
vent air into the container
50–60
mL
25–30 mL
Vented Micro-
Volume Inlet
H938
175
25
Small-volume vials
5–6 mL
2.5–3 mL
Micro-Volume Inlet,
with Large-Bore
Spikes
H938
751
25
Small-volume bags or bottles that
require a large-bore spike
5–6 mL
2.5–3 mL
Syringe Inlet
H938
176
25
60 mL Luer syringes (regardless of
the volume they contain)
5–6 mL
2.5–3 mL
OEM inlets are sterile, bio-compatible, non-pyrogenic, non-DEHP and contain no natural rubber
latex components.
NOTE:
The compounder will automatically use the highest available value for the standard
priming volume and half of that value for the minimum priming volume. However, you can
adjust these priming volumes in the Inlet Editor. For instructions, refer to Using the Inlet Editor
on Page 150.