2
Infusion Modes
2-20
COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL
071960491
Rate-Volume Infusions
This mode allows programming the flow rate and the infusion volume,
and the pump then runs at the programmed rate until the programmed
volume has been delivered. The pump then switches to the KVO (Keep
Vein Open) rate.
Rate-volume infusions can be programmed for primary or secondary
infusions. Secondary rate-volume infusions can be programmed without
stopping the primary infusion.
For instructions on programming rate-volume infusions, see
“Programming a Primary Rate-Volume Infusion,” 4-21 and
“Programming a Secondary Rate-Volume Infusion,” 4-46.
Volume-Time Infusions
This infusion mode lets the user enter the Volume to be Infused and
Time Duration, and the pump then calculates and displays the flow rate.
This feature is available in primary and secondary modes. Secondary
infusions can be programmed in Volume-Time without stopping the
primary infusion.
For instructions on programming Volume-Time infusions, see
“Programming a Primary Volume-Time Infusion,” 4-25 and
“Programming a Secondary Volume-Time Infusion,” 4-49.
COLLEAGUE GUARDIAN Feature
The COLLEAGUE GUARDIAN feature is a configurable option that
helps to reduce the potential for medication programming errors by
allowing program limits to be predefined for labels in the pump's label
library, including custom labels, based on facility or care area protocols.
The COLLEAGUE GUARDIAN feature allows the clinician to
compare pump programming with facility-defined dose guidelines at the
point of care. If the clinician programs values outside of the rule sets the
facility has programmed for a label, an out of limits warning occurs. The
COLLEAGUE GUARDIAN feature can be configured for rate-volume
and dose mode programming.
Note:
All of the setup required to use the COLLEAGUE GUARDIAN
feature must be completed by facility-authorized personnel
according to the instructions in the
COLLEAGUE Pump
Configuration Manual
or the instructions accompanying the
COLLEAGUE GUARDIAN Configuration Tool.
Summary of Contents for Colleague 3 CXE
Page 24: ...1 Indications for Use 1 12 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Page 50: ...2 Infusion Modes 2 26 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Page 54: ...3 Check out 3 4 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Page 152: ...7 Storage 7 8 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Page 228: ...Index 8 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Page 229: ...071960491 ...