11
Programming a Primary Rate-Volume Infusion
11-6
COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL
071960491
Programming a Primary Rate-Volume Infusion
1.
For single channel pumps: press the
Primary
soft key.
For triple channel pumps: press the desired
Channel Select
key.
The Main Display shows the Primary Rate-Volume
programming
screen, and the
Rate
field is highlighted (Figure 11-3).
2.
Enter the flow rate using the numeric keypad.
3.
Press the
Vol
key or use the
×Ø
keys to highlight the
Volume to be
infused (VTBI)
field.
4.
Enter the VTBI using the numeric keypad.
5.
Press the
Confirm Primary
soft key.
6.
Press
START
to start the infusion. The green RUNNING LED will
illuminate on the pump channel and a moving drop icon will appear
on the Main Display.
Confirm that flow is occurring by observing drops falling into the
drip chamber.
Note:
Pump Status
If the pump status is unclear, close any pop-up win-
dows on the display and press the
Main Display
key to
continue.
Figure 11-3 Rate-Volume
Programming Screen
! WARNING !
There may be periods of no flow for flow rates less than or equal to 1mL/hr.
! WARNING !
Do not enter a Volume to be infused greater than the amount of fluid available
in the container.
Summary of Contents for Colleague 3 CXE
Page 24: ...1 Indications for Use 1 12 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Page 50: ...2 Infusion Modes 2 26 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Page 54: ...3 Check out 3 4 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Page 152: ...7 Storage 7 8 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Page 228: ...Index 8 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Page 229: ...071960491 ...