Troubleshooting
Troubleshooting Alarms
Stuck Key Detected
STK KEY
(on all channels)
Without an infusion running, a
key on the pump’s front panel
has been pressed for longer than
5 seconds.
Do one of the following as appropriate:
• Release the key.
• Relocate any objects next to the pump that may
be pressing a key.
If unresolved for more than 50 seconds, a
failure will result.
Temperature Too High
TEMP HGH/label
Operating or administration set
temperature is outside the design
limits.
Move the pump, administration set, and
solution to a suitable temperature environment,
then restart the infusion.
Temperature Too Low
TEMP LOW/label
Operating or administration set
temperature is outside the design
limits.
• Move the pump, administration set, and
solution to a suitable temperature environment,
then restart the infusion.
• Allow cold solutions or sets to warm to the
operating temperatures before use.
X - Tube Loading in
Progress
LOADING
The administration set was not
fully loaded in the tubing
channel when the
START
key
was pressed.
Wait for the loading action to complete, then
press the
START
key.
X - Tube Misloaded
PATIENT---->>>>
• The administration set is
improperly loaded.
• The administration set was not
fully removed from the tubing
channel.
• A hardware problem may have
occurred.
•
Close the regulating clamp
on the
administration set
and remove the
administration set. See “Unloading the
Administration Set,” 4-70.
• If alarm occurs again, a hardware problem may
exist. Take the pump out of service and have it
inspected by Baxter-trained, qualified
personnel.
X - Tube Not Loaded
NO TUBE
The administration set was not
loaded prior to pressing the
START
key.
• Press the
Open
key to reset the alarm.
• Load the administration set, then press the
START
key
.
X - Unconfirmed Primary
Program
X - Unconfirmed
Secondary Program
STOPPED
The
START
key was pressed
prior to confirming
programming information.
Press the
Confirm Primary
or
Confirm
Secondary
soft key followed by the
START
key.
Note:
An
Unconfirmed Primary Program
alarm can occur when attempting to start a
secondary infusion if a primary infusion has been
programmed but not confirmed . To clear the
alarm, press the
Primary
soft key, then the
Confirm Primary
soft key. Press the
Secondary
soft key to continue with the
secondary infusion.
Table 8-1
Troubleshooting Alarm Messages — continued
Alarm Message
Cause
Recommended Action
Summary of Contents for Colleague 3 CXE
Page 24: ...1 Indications for Use 1 12 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Page 50: ...2 Infusion Modes 2 26 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Page 54: ...3 Check out 3 4 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Page 152: ...7 Storage 7 8 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Page 228: ...Index 8 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Page 229: ...071960491 ...