8
Troubleshooting Failures
8-6
COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL
071960491
1.
Press the appropriate
Channel Select
key. The parameters for the
infusion in progress when the channel failed are displayed (see
Figure 8-6 for an example). Record the parameters so the infusion
can be continued on one of the other pump channels if necessary.
2.
Perform Step 3 or Step 4 as appropriate:
3.
If the tube loading mechanism is open,
close the regulating clamp
on the administration set
and remove the set. Press the
Done
soft
key. The channel is now shown as
Out Of Service
.
Continue to step Step 5.
4.
If the channel fails with the tube loading mechanism in the “closed”
position:
Close the regulating clamp on the administration set.
Press the
Open
key for the failed pump channel. If the mechanism opens,
remove the set.
If the mechanism does not open (or if the device has been
powered off due to depleted batteries), use the Manual Tube
Release to remove the set (see “Using the Manual Tube Release,”
4-71).
5.
Press the
Done
soft key. The Main Display shows that the pump
channel is out of service (Figure 8-7).
6.
Allow any infusions running on the other pump channels to
complete. The failed pump channel cannot be used to deliver
infusions, but the
Volume
History
key can be used to retrieve history
information if desired.
7.
Remove the pump from service and have it inspected by
Baxter-trained, qualified personnel as soon as possible.
Figure 8-6 Channel Failure
Pop-up Message
Figure 8-7 Channel Failure
Display
Note:
Using Manual
Tube Release
If the Manual Tube Release is used following a channel
failure to remove the administration set, the pump can-
not be powered off until the Manual Tube Release has
been reset.
If three unsuccessful attempts are made to reset the
Manual Tube Release, a channel failure will occur and
the channel will be unavailable for use until the Manual
Tube Release is reset and the pump is powered off and
back on.
Summary of Contents for Colleague 3 CXE
Page 24: ...1 Indications for Use 1 12 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Page 50: ...2 Infusion Modes 2 26 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Page 54: ...3 Check out 3 4 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Page 152: ...7 Storage 7 8 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Page 228: ...Index 8 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Page 229: ...071960491 ...