Unloading the Administration Set
4-72
COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL
071960491
Resetting the Manual Tube Release
If a channel failure occurs and an attempt is made to power off the pump
without first resetting the Manual Tube Release, the Reset Manual Tube
Release pop-up is displayed (Figure 4-101).
Reset the Manual Tube Release as follows:
1.
Close the regulating clamp on the administration set.
Ensure
there is no administration set or foreign object in the tubing channel.
2.
Turn the release tab (Figure 4-100A) clockwise until it stops and
push the tab into its socket.
3.
For triple channel pumps: Repeat the steps above as needed for
additional channels.
4.
The
Done
soft key is displayed when the Manual Tube Release is
reset for all affected channels. Press the
Done
soft key to clear the
alarm.
Figure 4-101 Reset Manual Tube
Release Pop-up
Note:
MTR After
Channel Failure
If the Manual Tube Release is used following a chan-
nel failure to remove the administration set, the pump
cannot be powered off until the Manual Tube Release
has been reset. A Reset Manual Tube Release pop-up
message will be displayed.
Note:
Number of MTR
Attempts
If three unsuccessful attempts to reset the Manual
Tube Release are made, a channel failure occurs. The
pump cannot be used until the Manual Tube Release is
reset and the pump is powered off and back on.
Summary of Contents for Colleague 3 CXE
Page 24: ...1 Indications for Use 1 12 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Page 50: ...2 Infusion Modes 2 26 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Page 54: ...3 Check out 3 4 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Page 152: ...7 Storage 7 8 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Page 228: ...Index 8 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Page 229: ...071960491 ...