4
Stopping a Secondary Infusion Before Completion
4-68
COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL
071960491
To restart the infusion:
Stopping a Secondary Infusion Before Completion
1.
Press the
STOP
key on the desired pump module.
A stop icon displays on the Main Display (Figure 4-98) and the
RUNNING LED is no longer illuminated for that channel. The pump
module displays
STOPPED
.
2.
Close the On/Off clamp on the secondary administration set. If the
primary set has a regulating clamp above the pump, open the
regulating clamp.
3.
To start the primary infusion:
To restart the secondary infusion, close the regulating clamp on the
primary set (if applicable), open the On/Off clamp on the secondary
administration set, and:
For single channel pumps:
For triple channel pumps:
press the
Primary
soft key,
then press the
START
key.
press the
Channel Select
key
on the desired pump channel,
then press the
START
key.
Figure 4-98 Secondary Stopped
For single channel pumps:
For triple channel pumps:
press the
Primary
soft key,
then press the
START
key.
press the
Channel Select
key
on the desired pump channel,
then press the
Primary
soft
key, then press the
START
key.
For single channel pumps:
For triple channel pumps:
press the
Secondary
soft key,
then press the
START
key.
press the
Channel Select
key
on the desired pump channel,
then press the
START
key.
Note:
Do not cut the tubing to remove the administration set
from the channel. If the tubing is cut, remove the slide
clamp immediately.
Summary of Contents for Colleague 3 CXE
Page 24: ...1 Indications for Use 1 12 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Page 50: ...2 Infusion Modes 2 26 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Page 54: ...3 Check out 3 4 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Page 152: ...7 Storage 7 8 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Page 228: ...Index 8 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Page 229: ...071960491 ...