34
Setting Up the Pump
Manual 41018v0800
SIGMA Spectrum Infusion System
Revision B
Operator’s Manual
NOTE:
An icon is displayed next to any configuration item in the Options menu (User or
Biomed) where the setting has been assigned in the Drug Library. See Figure 12.
Display Settings
■
Run Screen Options.
Each of the items in this list may be set to On or Off. If enabled, they
are included in the alternating screens that are displayed while the Pump is running.
Audio level indicator – Shows L, M, or H (as selected in Audio Volume).
Rate mL/hr – Shows the mL/hr infusion rate of the current delivery.
Dose rate – Shows the dose rate of the current delivery.
mL VTBI – Shows the remaining Volume To Be Infused, in mL.
Time (hr:min) – Shows the time remaining in the current infusion, in hr:min format.
■
Display Adjust.
The display backlight brightness level may be adjusted from 1 to 10 (10 being
the highest) or set to Off. The backlight consumes approximately 400 mW when set to
maximum brightness; therefore, battery life is maximized when the backlight is set to Off.
To adjust display brightness:
1. Access the User Options menu screen:
From the Care Area selection screen, press the
options menu
soft key. Then use the
arrow soft keys to select
User Options
and press
OK
.
From the Infusion Running screen, press the
info settings
soft key.
2. Using the arrow soft keys, select
Display Settings
and press
OK
. The Display Settings menu
screen appears.
3. Using the arrow soft keys, select
Display Adjust
and press
OK
. The Display Adjust screen
appears.
4. Using the arrow soft keys, raise or lower the brightness to the desired level.
5. After selecting the desired brightness level, press
OK
to save the level and exit the Display
Adjust screen.
6. Press the
exit
soft key as needed to return to the Care Area selection screen or Infusion
Running screen.
View Information
■
Pump Information.
The Pump Information screen shows the following read-only information:
SW (software) Version
Serial number – The serial number assigned by Baxter for tracking and device history.
Figure 12. MDL Icon.