116
Alarms
Manual 41018v0800
SIGMA Spectrum Infusion System
Revision B
Operator’s Manual
Very Low Battery
If the battery level drops below the low-battery level, the message changes to Very Low Battery
and begins to flash. The back light also dims to reduce battery usage. See Figure 133.
NOTE:
When the Very Low Battery warning initiates, a minimum of 15 minutes of runtime
remains.
Battery Depleted
If the battery level drops below the Very Low Battery level, the message changes to Battery
Depleted. See Figure 134.
The Pump will stop running. If the Pump is not plugged in within 3 minutes, the Pump will shut itself
off.
Figure 133. Very Low Battery Alarm.
Figure 134. Battery Depleted Alarm.