Ensure that the plunger is fully retracted by 
pulling it to the back and that the cartridge 
is fully forward by pushing it forward. The 
inserter is now ready for assembly with the 
lens shuttle.

 

 

 

 

 

 

Remove the lid from the vial by peeling it 
away from you. Ensure that the Uchannel 
(highlighted in red) in the vial is facing 
towards you. Using the thumb and index 
finger to remove the lens shuttle from the 
vial.

Grasp the inserter body with one hand and 
the shuttle with the other hand. Ensure that 
the four legs of the shuttle are centered to 
the inserter body side walls and insert the 
shuttle straight down until you hear an 
audible click. Visually confirm that the 
shuttle is sitting flush and that the top of the 
shuttle is horizontal with all four legs inside 
the side walls of the inserter body.

Add a drop of viscoelastic* to the soft tip for easy 
entry into the shuttle.

 

 

1

 

 

6

While holding the inserter body with one 
hand, gently advance the plunger. 
Visually confirm that the soft tip enters 
the shuttle without deforming.

Continue to advance the plunger until the lens is in 
the conical section of the cartridge tip, as shown in 
the image. 
To deliver the lens, insert the cartridge tip into the 
incision with the tip bevel facing down. Slowly 
advance the plunger until the lens is fully released 
into the eye.

 

 

 

 

  

12

 

TAB

 

 

1

 

3

 

 

 

 

 

 

14

Open the box and remove the inserter and 
IOL pouch. Open the peel pouch to take the 
IOL vial out. Peel the tyvek lid and remove 
the inserter from the tray.

VISCOPORT

Hold down the tab on the top of the shuttle and 
apply viscoelastic* into the viscoport of the 

 

cartridge. Visually verify that the viscoelastic has 
traveled up to the mark as shown in the image.

 

 

 

1

 

 

 

5

Soft Tip

7

8

Conical Section

 

Tip

Scan the QR code to watch the loading video

Loading Instructions 

Akreos

®

 AO and Akreos® AO MICS preloaded with the Lomb 

SimplifEYE™ delivery system