30
VLXN
Camera Type
Temperatures for emergency shutdown
(at internal temperature sensor)
Monochrome
VLXN-650M.I.JP
69 °C (156.2 °F)
4.3 Mechanical Tests
except .EF cameras
Environmental
Testing
Standard
Parameter
Vibration,
sinussodial
IEC 60068-2-6
Continuous
oscillation
10-2000 Hz
Amplitude un-
derneath cross-
over frequencies
0,75 mm
Acceleration
1 g
Test duration
150 min (axis)
450 min (total)
Vibration, broad
band
IEC 60068-2-64
Frequency
range
10-2000 Hz
Acceleration
10 g
Test duration
5 h (axis)
15 h (total)
Shock
IEC 60068-2-27
Pulse time
11 ms / 6 ms
Acceleration
50 g / 100 g
Bump
IEC60068-2-29
Pulse Time
2 ms
Acceleration
100 g
.EF cameras only
Environmental
Testing
Standard
Parameter
Vibration, broad
band
IEC 60068-2-64
Frequency
range
5-150 Hz
Acceleration
0.5 g
Test duration
5 h (axis)
15 h (total)
Shock
IEC 60068-2-27
Pulse time
30 ms
Acceleration
5 g