background image

Issue 2: 2016          

                 SERVICE & GUARANTEE

 

 

8.2  How to make a Guarantee Claim

 

 

On  receipt  of  the  goods  and  order  the  Company  will 

examine  the  goods  and  determine  the  nature  and 

cause  of  the  defect.  If  the  defect  is  covered  by  the 

guarantee  a  replacement  or  repair  will  be  effected  at 

the Company's expense.  If the defect is not covered by 

the  guarantee  the  Company  will  quote  the  Purchaser 

for the replacement or repair, and will not proceed until 

written acceptance of such quotation is received. 

 

During the first year of the guarantee the Company will 

bear  the  cost  of  returning  guarantee  replacement  or 

repaired  goods  to  the  purchaser's  premises  in  the 

country  to  which  the  goods  were  originally  delivered.  

At the purchaser's request and expense, the Company 

will return goods to other countries. 

 

During the second to the fifth year of the guarantee the 

return of guarantee replacement or repaired goods 

shall be at the Purchaser's expense. 

 

Always adhere to safe local regulatory disposal practise 

for alkaline batteries used within the AUDIT Instrument. 

 

Care should be taken that the couplant used is neither 

harmful to the operator nor will it attack the probe.  For 

advice please contact Baugh & Weedon. 

 

 

Page 18 

 

 

Issue 2: 2016                          

  

OPTIONAL EQUIPMENT

 

                                                

 

1.3   Optional Equipment 

   

 

 

 

 

         Item 

Transducer  type 2.5MHz  BD52  DT35 5MHz/0.25” 

Transducer  type 7.5MHz  BD17  DT37 7.5MHz/0.87” 

Transducer  type 5.0MHz  BD25  DT39 Hi-Temp              

 

  

2.    TECHNICAL SPECIFICATION

 

 

2.1  

Thickness Range Interface Echo technique 

 

Minimum     0.63mm   Maximum    500.0mm  

 

0.025” 

19.0 in 

 

2.2  Accuracy & Calibration

 

 

Velocity  

Range 1200m/s  to 10000ms   

 

0.0492 in/µs to  0.3930in/µs 

  

Resolution:  0.01mm. (0.001” 

Accuracy:    

± 0.01millimeters ±0.001inches     

 

   

 

Material dependent 

 

2.3  Power Supply

 

 

Source:   2 x LR6( AA)  Alkaline Batteries  or 

rechargeable equivalents 

Battery life:  50 hours continuous (alkaline dry batteries) 

Auto power off timer: 5 Minutes

.

  

 

Page  4 

Issue 2: 2016      

          

INTRODUCTION 

 

Before  using  the  equipment,  operators  should  make 

themselves  fully  conversant  with  the  operating 

procedures as laid out in this handbook 

 
1.2   Standard Schedule of Equipment 

 

Dimension 121 x 70.3 x 32.3mm 

Weight 295g including batteries  

 

The  standard  equipment  package  consists  of  the 

following.  

 

Product 

                                                    

  

 

 

 

AUDIT 207 Thickness Gauge.  

0704/A207 

  

 

Complete with two AA alkaline batteries   

 

  Plastic Transit/Carrying Case 

0704/0175. 

 

 

 

  Probe type 5MHz   BD25 

DT35 5MHz 

 

 

 

 

  Operator Handbook. 

0704/0201 

 

 

 

 

 

  Ultrasonic Coupling Agent 

0704/185 

 

 

 

 

 

 

 

 

Page  3 

 

 

Issue 2: 2016

                              

SERVICE & GUARANTEE

 

              

   

8.   SERVICE & GUARANTEE OF EQUIPMENT 

 

8.1  Guarantee 

 

AUDIT  207  is  guaranteed  for five  (5)  years,  excluding 

batteries and transducers  which have been subject to 

wear. 

 

Baugh & Weedon ("the company") undertakes to make 

good  by  sending  a  replacement  or,  at  it's  option,  by 

repairing defects arising from faulty design, material or 

workmanship within a five (5) year period from the date 

of dispatch.  

 

Your existing consumer rights are not affected by these 

guarantees and conditions. 

 

Always adhere to site rules and guidelines concerning 

Health & Safety when using the AUDIT 207 Instrument. 

 

Always wear safety equipment appropriate to the area 

where the measurement is to be made. 

There  are  no  user  serviceable  parts  inside  the 

instrument case of the AUDIT 207  Service work should 

only be carried out by qualified personnel  . 

 

 

 

 

Page 17 

 

Summary of Contents for Audit 207

Page 1: ...4 2 TECHNICAL SPECIFICATION 4 2 1 Thickness Range 4 2 2 Accuracy Calibration 4 2 3 Power Supply 4 2 4 Temperature Range 5 2 5 Environmental Protection 5 2 6 Electromagnetic Compatibility 5 3 PRINCIPLE OF OPERATION 6 4 DIRECTIONS FOR USE 7 4 1 Preparation 7 4 2 Transducer ZERU Calibration 7 4 3 Thickness Measurement 8 4 4 Two point Calibration 8 4 5 Metric Imperial Mode change 9 4 6 Scan Mode 9 4 7...

Page 2: ...acy Calibration Velocity Range 1200m s to 10000ms 0 0492 in µs to 0 3930in µs Resolution 0 01mm 0 001 Accuracy 0 01millimeters 0 001inches Material dependent 2 3 Power Supply Source 2 x LR6 AA Alkaline Batteries or rechargeable equivalents Battery life 50 hours continuous alkaline dry batteries Auto power off timer 5 Minutes Page 4 Issue 2 2016 INTRODUCTION Before using the equipment operators sho...

Page 3: ... reflected back from the far surface As the velocity at which the sound travels is constant for a given material the thickness can be computed and displayed Thickness readings are held for 5 Minutes after the probe is removed from the work piece Page 6 Issue 2 2016 TECHNICAL SPECIFICATION Backlight is operated by the UNIT key selectable on off Low battery is indicated by the battery symbol located...

Page 4: ...highest known thickness of the test part 2 Press CAL key until set thickness is displayed using up down keys to display the known thickness Page 8 Issue 2 2016 DIRECTION FOR USE The velocity used for calculation is that of steel It is a simple matter to convert the reading for other common materials by using the appropriate Velocity Through the CAL key operation Fit batteries see section 4 4 Conne...

Page 5: ...ce of the material under test it is possible due to the composition and or grain structure of some materials plastics cast iron etc that limitations to the thickness range is possible as the material is too attenuative for the ultrasound to penetrate Page 10 Issue 2 2016 PREPARATION 3 Press ZERO key 4 Place the probe on the thinner known thickness of the test part followed by the CAL key 2 nd thic...

Page 6: ... adjust up down keys to display the known thickness 3 Press ZERO key 4 Place the probe on thinner known section followed by the CAL key 2 nd thickness is displayed using up down keys set this thickness displayed to match thinner section press CAL key Velocity will automatically enter measurement screen The Audit 207 is now ready to take readings AUDIT 207 Two Point Calibration 1 Place the probe on...

Page 7: ... incorporated on 13 11 2006 in the United Kingdom Company Registration Number 05996485 VAT Number GB583598190 ALPINE C O M P O N E N T S Postal Address Alpine Components Ltd Innovation Centre Highfield Drive Churchfields St Leonards on Sea TN38 9UH United Kingdom Telephone 01424 858118 Website Address www alpine components co uk ...