
MVT26
October 2003
Section 4
Page 73
4.1.3
Chamber Safety Bore
The chambers in all BAUER purification systems are designed to prevent pressurization if the cartridge is
missing, not seated properly or damaged (See Figure 4.1-1). Without a cartridge properly in place the
safety bore is not sealed, the air escapes into the atmosphere, no pressure can be built up and thus it is
ensured that unfiltered air is not supplied to the consuming device. If air is escaping from the safety bore
remove and check cartridge. If necessary replace the cartridge or O-rings.
4.1.4
Manual Condensate Drainage
The condensate must be drained from the oil and water separator before changing any cartridge, before
beginning each filling procedure and in the absence of an Automatic Condensate Drain System, every fif-
teen minutes during the filling procedure. This is done by slowly opening the condensate drain valves.
They are opened approximately 1/3 of a turn to the left and held open until the condensate is completely
drained. The condensate drain valves close by spring pressure but if necessary may be tightened by hand
to ensure they are completely air tight.
4.1.5
Model, Serial Number and Part Number Identification
4.1.5.1
Compressor Dataplate
The model number, date of manufacture and serial number can be found of the compressor unit identifi-
cation plate on the compressor unit’s frame.
4.1.5.2
Purification System Dataplate
Refer to the compressor unit purification system dataplate (See Figure 4.1-3) on the compressor front to
determine your purification system model and specifications.
Figure 4.1-1
Cartridge Safety Venting
Figure 4.1-2
Compressor Unit Identification Plate
Cartridge Installed
No Cartridge Installed
Venting
Bore
BAUER
COMPRESSORS
BAUER COMPRESSORS, INC.
NORFOLK, VIRGINIA U.S.A
LBL-004
MODEL NO.
BLOCK NO.
CAPACITY
CFM
CHG. RATE
CFM
MOTOR
HP
SPEED
RPM
SERIAL NO.
PRESSURE
PSIG
VOLTS
PH
HZ
AMPS
DATE OF MANUFACTURE
DATE OF MANUFACTURE