Operation and Features
3-6
BASLER A102
k
DRAFT
Case 2 - Exposure Start With Frame Valid Low
Figure 3-5: ExSync, Level-controlled Mode - Exposure Start with Frame Valid Low
If the ExSync signal falls while frame valid is low as shown in Figure 3-5:
• Exposure will start after a delay of 12.4 µs.
• The actual length of the exposure time will be equal to the ExSync signal low time plus 15 µs.
L
As shown in Figure 3-5, FVAL
must
be low for at least 1 µs before the ExSync signal
falls.
Timing
charts are
not drawn to
scale.