Electromagnetic Flowmeter
BF
M921
BASI
22
Operation manual
We recommended using software FlowAssistant for easy Extended Calibration. It contains „calibration
wizard“ and can prevent flowmeter from incorrect calibration.
Extended Calibration is available via the USB interface only.
User can change calibration values only
if the correct calibration password has been entered. Default factory
set Calibration password is “10000”.
4.3
Nominal values (standard factory setting)
4.3.1
General
General
Language
English
Display/Backlight
On
Display/Contrast
Medium
Display/Startup
Overview
Display/Aux item
Aux. Volume
Date & time/Time
Actual time
Date & time/Date
Actual date
Date & time/Format
M/D/Y
Basic password
00000
Calibration password
10000
Table 8 : Nominal values - general
4.3.2
Measurements
Flow
Unit
m3/h
Resolution
default
User constant
1.0
Direction
Positive
LF cutoff
flowrate Q 1% /2
MA samples
4 s
Empty pipe detect
On
Volume
Unit
m3
Resolution
default
User constant
1.0
Datalogger
Interval
Off
Temperature
Setting any new value in calibration menu changes calibration data! Calibration
should be performed in an appropriate equipped laboratory.
User can restore calibration settings to factory values
(Setup menu/General/
Factory settings/
Restore calibration points).