background image

17 

 

3.

 

CALIBRATION 

Calibration must be carried out on a new sensor the next time the instrument is turned on; 

with an already installed sensor, the calibration should be carried out at specified intervals. The 
A-CAL firmware calibrates the Gas-Rover using calibration gas available from Bascom-Turner 
containing 2.5 + 0.05% methane and 100 ppm + 2 ppm CO in air. Calibration gas is provided in 
a disposable tank containing either 105 liters of gas (Part No. MC-105) or 620 liters (part No. MC- 
620). The thermal conductivity sensor is automatically calibrated for 100%  methane during the 
autozero  routine  preceding  calibration  with  the  One-Gas  calibration  algorithm.  It  can  be  also 
calibrated  with  pure  methane  or  system  gas.  The  oxygen  sensor  is  calibrated  with  ambient  air 
during power-up. 

The calibration gas must be delivered at or near atmospheric pressure to ensure accurate 

calibration.  For  the  105  liter  gas  tanks  Bascom-Turner  recommends  using  a  constant  flow 
regulator (Part No. FR-001) set to the sampling rate of the Gas-Rover. For the 620 liter gas tanks 
Bascom-Turner  recommends  using  an  on-demand  regulator  (Part  No.  FR-401),  which 
automatically adjusts its flow to the pumping rate of the Gas-Rover. 

A-CAL  Operation. 

Calibrate  the  thermal  conductivity  sensor  with  methane  or  natural 

gas first, and then calibrate the CO and catalytic combustion sensors with 2.5% methane and 100 
ppm  CO  in  air.  Alternatively,  use  the  One-Gas  calibration  algorithm  and  calibrate  the  CO  and 
catalytic combustion sensors with 2.5% methane and 100 ppm CO in air. 

A-CAL Calibration with Pure Methane or System Gas. 

Power on the Gas-Rover with 

the left “Power-On” button and while the unit displays “

Calibrate By …..

” press the “CAL button. 

After autozero, choose “A-CAL”, connect pure methane or natural gas and press the left “Yes” 
button. The screen displays “

Calibrating

” for about a minute and then changes to “

Calibration 

OK

” with a display of readings taken while sampling calibration gas or “

Does Not Calibrate

” if 

calibration was unsuccessful. Three tries are allowed before the screen displays “Please Service” 
for 15 seconds and then the instrument turns OFF. 

Calibration with  2.5% Methane and 100 ppm CO. 

Power on the Gas-Rover with the 

left  “Power-On”  button  and  while  the  unit  displays  “

Calibrate  By  …..

”  press  the  right  “CAL” 

button  once.  After  a  successful  autozero,  choose  A-CAL.  With  the  unit’s  screen  displaying 

Connect Calibration Gas

”, connect the 2.5% methane and 100 ppm CO in air calibration gas and 

press the left “Yes” button.  The screen displays “

Calibrating

” for about a minute and then changes 

to “

Calibration OK

” with a display of % GAS (and ppm CO readings if it has a CO sensor) or 

Does Not Calibrate

”, if calibration was unsuccessful. Pressing the left “Re-Try” button returns 

to the Calibration screen. If three tries at calibration are unsuccessful, the screen displays “

Please 

Service

” for 15 seconds and then the instrument turns OFF. 

Following  calibration,  an  intermediate  screen  displays  the  sensitivity  in  µw/ppm  and 

mw/1%.  The  first  quantity  (µw/ppm)  refers  to  the  Survey  mode,  the  second  (mw/1%)  to  the 
Monitor  mode.  If  the  sensitivity  of  the  Survey  mode  falls  below  0.5  µw/ppm,  the  natural  gas 
sensor should be changed. Similarly if the sensitivity in the Monitor mode falls below 8 mw/1% 
GAS, the gas sensor should be changed. In both cases the display reads 

“Please Change Methane 

Sensor” 

and the instrument is placed in Off condition until a new sensor is installed. 

PART 3. THE USB INTERFACE 

CAUTION: 

Automatic  calibration  presupposes  and  depends  on  using  Bascom-Turner’s 

calibration gas (Part No. MC-105 or MC-620). Do not use a gas of a different 
composition for automatic calibration. 

Summary of Contents for GAS-ROVER VGI-201

Page 1: ...BASCOM TURNER INSTRUMENTS GAS ROVER DETECTORS OPERATION MANUAL NATURAL GAS CARBON MONOXIDE AND OXYGEN DETECTORS Part Number OM 0717...

Page 2: ...s must be trained to recognize instrument malfunctions and to have such instruments repaired promptly Bascom Turner offers no warranty that failures will not occur only that when they occur they will...

Page 3: ...s Liquids or RFI 9 6 Change of Batteries 10 7 Probes 11 8 Routine Maintenance 12 Appendix 1 Main Key Functions 14 Appendix 2 The Gas Rover in Mobile Surveys 14 PART 2 SETUP AND CALIBRATION 1 Setup 15...

Page 4: ...after gas is pumped for a precise fixed time Automatic Data Storage Automatic storage of time and mode stamped data readings Automatic storage the last 24 calibrations Automatic generation of reports...

Page 5: ...Time 60 seconds Response Time CH4 0 6 seconds Survey 2 seconds Monitor CO 45 seconds O2 60 seconds Operating Temp 20o C to 40o C 0o F to 105o F Storage Temperature 40o C to 60o C 40o F to 140o F Data...

Page 6: ...pes and other conduits All models which detect carbon monoxide CO may be used to test for CO in ambient air flue gas and gases given off by appliances A detector alerts the user acoustically and visua...

Page 7: ...ion without this filter will eventually degrade the sensors and the pump It also voids the limited warranty It is essential that the filter be replaced monthly or sooner if it turns brown Operation wi...

Page 8: ...nt gas refer only to the calibration gas and can be relied upon only in this respect in assessing an atmosphere sampled by the detector Furthermore concentrations displayed by the detector refer to a...

Page 9: ...g off of power The next display suggests a Calibrate by a date thirty days from the date of the last calibration and an opportunity to calibrate see Part 2 Calibration If calibration is not chosen the...

Page 10: ...shows the concentration of gas in air in parts per million by volume ppm The beeper will sound if the ambient concentration of gas rises above a user selected value see Part 2 Section 1 The factory se...

Page 11: ...the front panel to Bar Hole BH The instrument automatically clears any residual gas and checks for clean air The pump is then turned off and the instrument is ready for bar holing After inserting the...

Page 12: ...ert the probe into the flue and read the concentration of CO in parts per million A flue gas check of CO should take one minute or less For every minute of flue gas sampling allow the detector a minut...

Page 13: ...tomatic Sensor Check If a gas sensor fails opens up the display shows Replace Methane Sensor Automatic Pump Check If the intake is blocked the display shows Flow Blocked and the detector beeps until t...

Page 14: ...on monoxide Models VGO 321 and VGO 322 methane and oxygen detectors use the methane sensors listed above and iv A two electrode electrochemical cell for oxygen Gases or liquids with appreciable vapor...

Page 15: ...rms and the display flashes Low Batteries between readings If battery life is over the unit displays Low Batteries and powers down the batteries must be changed to make the instrument operational agai...

Page 16: ...fitting Flue Gas Probe The flue gas probe Part No FP 110 is a telescoping metal probe screwed finger tight into the flue gas filter The other end of the filter attaches to the dust and water block fi...

Page 17: ...block filters as they will not block water or remove very fine dust and dirt If this filter gets blocked it should be replaced Water stopper The water stopper Figure 2 provides a quick way to continue...

Page 18: ...th fresh filter material up to the bottom of threads place the plastic retainer disk flat on top and screw the top finger tight Check the seal by carrying out a block test with the detector Heavy Hydr...

Page 19: ...urvey kit Part No MP 001 which can be installed in about 4 hours MODIFICATIONS OF EXISTING SAMPLING SYSTEMS The two chain and funnel assemblies that run from the front bumper should be about four feet...

Page 20: ...ted with the right Update button When a flashing parameter is OK press the left Next button to either advance to the next digit or to leave the Edit screen and display the edited parameter Alarm Limit...

Page 21: ...Typically the pump current rises by about 15 ma when a hose and probe are attached The Pump Noise test follows when the Exit key is pressed Pump Noise is meaningful only under load i e when hose probe...

Page 22: ...o choose A CAL connect pure methane or natural gas and press the left Yes button The screen displays Calibrating for about a minute and then changes to Calibration OK with a display of readings taken...

Page 23: ...ading setup parameters DATA LINK can be installed on PC s with Windows 2000 or XP Pro operating systems and equipped with a USB port Figure 3 Main Menu the initial DATA LINK screen 1 DOWNLOADS The mai...

Page 24: ...ble which helps to identify when the unit was powered on and the operational mode changed Using the data on alarm flags in the Readings Summary file a Sorted Alarm Events table can be displayed to ide...

Page 25: ...of the data is archived in the PC no matter how many times it is downloaded from the unit to the PC Downloading Readings Data It can take 5 minutes to download all the readings data from a unit to a P...

Page 26: ...story by Unit ID is a report of the date and time of each calibration with before and after calibration readings of each sensor Since the pump must work in order to calibrate a detector this is checke...

Page 27: ...y User is a list of each time when a user is assigned a new unit The Unit Assignment History by User report summarizes the assignment audit trail with a chronological list of units used by a user and...

Page 28: ...23 Readings and Exposure Report Selection Form Bar Holing reports showing peak and sustained readings...

Page 29: ...back of the case see Step 2 above removing the manifold from the bottom of the case and unscrewing the four screws which secure the pump gasket pump plate diaphragm and piston to the manifold Usually...

Page 30: ...ase and main circuit board removed CAUTION After replacing a methane or a CO sensor perform the New Filament procedure described under Part 2 Setup and Calibration and then completely calibrate the in...

Page 31: ...ter 5 pkg FF 105 Heavy Hydrocarbon Filter 5 pkg HF 005 Inlet Filter Quick Connect Air Intake 5 pkg IF 705 3 Sensors Methane Sensor for Rovers with S N less than 401000 MS 601 Methane Sensor for Rovers...

Page 32: ...27 Bascom Turner Instruments 111 Downey Street Norwood MA USA 02062 781 769 9660 800 225 3298 Fax 781 769 2099 bascomturner com...

Reviews: