Installing the LC-23C in a CC-5 Compartment
1.
Lower the CC-5 lid and remove the utility cover from the back panel of the CC-5 cabinet
(four screws).
2.
On some models, pressed-in, threaded nuts will be located on the outside of the back
panel (Figure 6). These must be removed. DO NOT REMOVE IF THEY ARE ON THE
INSIDE OF THE CABINET. To remove external nuts, screw a long machine screw (6-32
thread) into the nut from the rear until it protrudes through the nut. Use pliers for
leverage and gently work the screw from side to side until the nut comes loose.
3.
Remove the LC-23C cover and set aside. Push the heater block through the now open
utility access in the CC-5 back panel. On some models, the two registration posts on
the CC-5 inner panel will engage the LC-23C block for support.
4.
Secure the LC-23C to the cabinet using four 6-32 x 1/2" machine screws and four 6-32
lock nuts. The lock nuts are not necessary if pressed-in nuts are available on the inside of
the CC-5 cabinet. The mounting plate should now be flush with the outside of the back
panel on the CC-5 cabinet.
Figure 6.
Installation of the LC-23C in a BAS 480 or CC-5 Electrochemical Detector. (A) Location of four screws (B) Breaking off
the pressed-in nuts (C) Left Side View of Attachment Hardware.
6-32 LOCKNUT
(4 PLACES)
6-32 x 1/2" SCREW
(4 PLACES)
SCREW
LOCATIONS
A
B.
C.
CC-5
REAR
CC-5
FRONT
Section 3. INSTALLATION
LC-23C COLUMN HEATING COMPARTMENT
6