OPERATION (cont.)

USING THE 5-HOLE DIAPHRAGM

1. To obtain the best contrast for observing, match the hole size to  

    the objective that is being used to view the specimen. 

2. Each hole has a corresponding number from 1 to 5. 1 is the    

    smallest hole; 5 is the largest hole.

    Use the following guidelines to match the hole number to the  

    objective that you have selected: 

    40x objective: Use #5 hole

    10x objective: Use #4 or #3 hole

      4x objective: Use #2 or #1 hole

COARSE KNOB ADJUSTMENT - Model AY11240
1. The coarse adjustment knob has an adjustable heavy-light nut  

    (See Fig.1).

2. To adjust the knob loosen or tighten the nut.

    NOTE: Adjusting the nut too tight will make focusing difficult.  

    Adjusting the nut too loose will cause the tube to slide.  

Heavy-Light

Adjustment Nut

Fig. 1- Coarse Adjustment Knob

7. To clearly see the outline of the   

    specimen, rotate the coarse  

 

    adjustment knob and lower 

    the barrel to the space limiter.

8. Rotate the fine adjustment knob     

    until the image is in sharp focus.  

    When using other objectives, rotate  

    the fine focus adjustment until the  

    image is in focus.

6. To clearly see the outline of the   

    specimen, rotate the coarse  

 

    adjustment knob and lower 

    the barrel to the space limiter.

7. Rotate the fine adjustment knob     

    until the image is in sharp focus.  

    When using other objectives, rotate  

    the fine focus adjustment until the  

    image is in focus.

Model AY11240

Model AY11238

6

MODEL AY11228/AY11232

MICROSCOPE USAGE

BARSKA Model AY11228 and Model AY11232 are designed for 

biological studies such as specimen examination. They can also   

be used for examining bacteria and for general clinical and medical 

studies. Simple design and use is especially useful for school 

classroom instruction.

CONSTRUCTION

BARSKA Model AY11228 is a fixed power stereo microscope. It is 

constructed with two optical paths at the same angle. It is 

equipped with transmitted illumination and oblique illumination.  

By using this instrument, the user can observe and enlarge the 

right side stereo image. BARSKA Model AY11232 is a zoom stereo 

microscope. The object being viewed is enlarged through two 

identical sized sets of right and left eye lenses. The zoom provides 

different magnification and features an inversion system which 

allows the image to be viewed normally and right side up.

Model AY11228

Model AY11232

Diopter
Adjustment

Diopter
Adjustment

Prism
Cap

Prism
Cap

Eyepiece

Stage

Lens

Stage

Vertical 

Pole

Illumination

Controls

Rotary

Case

Illumination

Controls

Spring 

Clips

Spring 

Clips

Focus

Knob

Focus

Knob

Magnification

Adjustment

Knob

Focus

Knob

Tightening

Knob

Oblique

Illuminator

Oblique

Illuminator

Lens 

Housing

Eyepiece

Summary of Contents for AY11240

Page 1: ...ify the customer accordingly BARSKA reserves the right to replace any product which has been discontinued from its product line with a new product of comparable value and function This warranty shall...

Page 2: ...ther 1 2 Dry Weather 1 1 4 After use cover the instrument with the plastic dust cover 5 If instrument is to be stored for an extended period of time remove the eyepiece and oculars and store in a mois...

Page 3: ...parts before assembling 2 Attach 4x 10x and 40x objectives to revolving turret 3 Place the specimen on the stage and secure with spring clips NOTE The cover glass must face upward the thinner glass is...

Page 4: ...stment knob until the image is in sharp focus When using other objectives rotate the fine focus adjustment until the image is in focus Model AY11240 Model AY11238 6 MODEL AY11228 AY11232 MICROSCOPE US...

Page 5: ...TION 1 Remove components from package identify all parts before assembling 2 Tighten the knob on the stand to prevent the elevator from sliding down 3 Fix the binocular body on the stand with the tigh...

Page 6: ...ft eyepiece and bring a specific point into focus using the focus knob b By turning the diopter ring adjustment for the left eyepiece bring the same point into sharp focus c Then bring the same point...

Page 7: ...ATION 1 Remove components from package identify all parts before assembling 2 Tighten the knob on the stand to prevent the elevator from sliding down 3 Fix the binocular body on the stand with the tig...

Page 8: ...both eyes at the same time the diopter ring may need adjustment DIOPTER RING ADJUSTMENT 1 To adjust the eyepiece for viewing with or without eyeglasses and for differences in acuity between the right...

Page 9: ...e 16 PARTS LIST OPERATION 1 Remove all components from package Identify all parts before assembling instrument 2 Attach 4x 10x and 40x objectives by screwing into revolving turret Tighten and secure t...

Page 10: ...wiped off all of the cedar oil OPERATION cont ADJUSTING THE CONDENSER APERTURE 1 The numerical aperture of the condenser should match the numerical aperture of the objective being used 2 To make sure...

Reviews: